sauermann KIMO Instruments CP 116 Quick Start Manual Download Page 2

English

Quick Start Guide

Output / Supply

Active sensor 0-10 V or 4-20 mA (24 Vac/Vdc power supply ±10%), 3-4 wires

Passive loop 4-20 mA (power supply 16/30 Vdc), 2 wires

Common mode voltage <30 VAC

Maximum load: 500 Ω (4-20 mA) / minimum load: 1 K Ω (0-10 V)

Consumption

2 VA (0-10 V) or 0.6 VA (4-20 mA)

European directives

2014/30/EU EMC; 2014/35/EU Low Voltage; 2011/65/EU RoHS II; 2012/19/EU WEEE

Electrical connection

Screw terminal block for cables from 0.05 to 2.5 mm

2

 or from 30 to 14 AWG 

Carried out according to the code of good practice

PC Communication

USB-mini DIN cable

Environment

Air and neutral gases

Response time

< 10 seconds

Type of fluid

Air and neutral gases

Tolerated overpressure

4136 hPa

Conditions of use 
(°C/%RH/m)

From 0 to +50 °C. In non-condensing condition. From 0 to 2000 m

Storage temperature

From -10 to +70 °C

Atmospheric pressure 
transmitter

Features

• 

0-10 V or 4-20 mA output, active, power supply 
24 Vac/Vdc (3-4 wires) or 4-20 mA output, passive 
loop, power supply from 16 to 30 Vdc (2 wires)

• 

ABS V0 housing, IP65, with or without display

• 

“¼ turn” system mounting with wall-mount plate

• 

Housing with simplified mounting system

Indication of the pressure 
in mbar, mmHG or hPa

Measuring range from
800 to 1100 hPa

General features

Material

ABS V0 as per UL94

Protection

IP65

Display

LCD 10 digits.

Dimensions: 50 x 17 mm

Height of digits: values: 10 mm; units: 5 mm

Connections

Ribbed, Ø 6.2 mm

Cable gland

For cables Ø 8 mm maximum

Weight

143 g

Features of the housing

Symbols used

For your safety and in order to avoid any damage of the device, please follow the procedure described in 
this document and read carefully the notes preceded by the following symbol: 
 
The following symbol will also be used in this document, please read carefully the information notes 
indicated after this symbol: 

Connections

Inside the front housing

1

1.  Inactive switch
2.  Active switch
3.  Pressure connection 

4.  LCC-S connection
5.  Output terminal 

block

6.  Power supply 

terminal block

7.  Cable gland

Removable front face

Fixed back housing

4

7

6

5

2

3

Summary of Contents for KIMO Instruments CP 116

Page 1: ...CP 116 Quick Start Guide EN FR www sauermanngroup com ...

Page 2: ...c Vdc 3 4 wires or 4 20 mA output passive loop power supply from 16 to 30Vdc 2 wires ABSV0 housing IP65 with or without display turn system mounting with wall mount plate Housing with simplified mounting system Indication of the pressure in mbar mmHG or hPa Measuring range from 800 to 1100 hPa General features Material ABS V0 as per UL94 Protection IP65 Display LCD 10 digits Dimensions 50 x 17 mm ...

Page 3: ...nergized Then you can make the settings required with the DIP switches as shown on the drawing below When the transmitter is configured you can power it up To configure the transmitter unscrew the 4 screws from the housing then open it DIP switches allowing the different settings are then accessible 1 2 3 4 Active switch Output setting Units setting 1 2 3 4 Active switch Please refer to the user m...

Page 4: ... Temps de réponse 10 secondes Type de fluide Air et gaz neutre Surpression admissible 4136 hPa Conditions d utilisation C HR m De 0 à 50 C En condition de non condensation De 0 à 2000 m Température de stockage De 10 à 70 C Capteur transmetteur de pression atmosphérique Caractéristiques Sortie 0 10V ou 4 20 mA active alimentation 24Vac Vdc 3 4 fils ou sortie 4 20 mA boucle passive alimentation de 1...

Page 5: ... capteur le mettre hors tension puis procéder aux réglages souhaités en disposant les interrupteurs comme décrit ci dessous Remettre le capteur sous tension une fois les réglages effectués Pour configurer le capteur dévisser les 4 vis du boîtier puis l ouvrir Le switch permettant les différents réglages est accessible 1 2 3 4 Switch actif Réglage de la sortie Réglage des unités 1 2 3 4 Switch acti...

Page 6: ......

Page 7: ...com Download the CP 116 data sheet Télécharger la fiche technique du CP 116 Download the LCC S software user manual Télécharger la notice d utilisation du logiciel LCC S https sauermann en custhelp com Customer service portal Portail service clients Utilisez notre Portail service clients pour nous contacter Use our Customer service portal to contact us ...

Reviews: