English
Quick Start Guide
Output / Supply
Active sensor 0-10 V or 4-20 mA (24 Vac/Vdc power supply ±10%), 3-4 wires
Passive loop 4-20 mA (power supply 16/30 Vdc), 2 wires
Common mode voltage <30 VAC
Maximum load: 500 Ω (4-20 mA) / minimum load: 1 K Ω (0-10 V)
Consumption
2 VA (0-10 V) or 0.6 VA (4-20 mA)
European directives
2014/30/EU EMC; 2014/35/EU Low Voltage; 2011/65/EU RoHS II; 2012/19/EU WEEE
Electrical connection
Screw terminal block for cables from 0.05 to 2.5 mm
2
or from 30 to 14 AWG
Carried out according to the code of good practice
PC Communication
USB-mini DIN cable
Environment
Air and neutral gases
Response time
< 10 seconds
Type of fluid
Air and neutral gases
Tolerated overpressure
4136 hPa
Conditions of use
(°C/%RH/m)
From 0 to +50 °C. In non-condensing condition. From 0 to 2000 m
Storage temperature
From -10 to +70 °C
Atmospheric pressure
transmitter
Features
•
0-10 V or 4-20 mA output, active, power supply
24 Vac/Vdc (3-4 wires) or 4-20 mA output, passive
loop, power supply from 16 to 30 Vdc (2 wires)
•
ABS V0 housing, IP65, with or without display
•
“¼ turn” system mounting with wall-mount plate
•
Housing with simplified mounting system
Indication of the pressure
in mbar, mmHG or hPa
Measuring range from
800 to 1100 hPa
General features
Material
ABS V0 as per UL94
Protection
IP65
Display
LCD 10 digits.
Dimensions: 50 x 17 mm
Height of digits: values: 10 mm; units: 5 mm
Connections
Ribbed, Ø 6.2 mm
Cable gland
For cables Ø 8 mm maximum
Weight
143 g
Features of the housing
Symbols used
For your safety and in order to avoid any damage of the device, please follow the procedure described in
this document and read carefully the notes preceded by the following symbol:
The following symbol will also be used in this document, please read carefully the information notes
indicated after this symbol:
Connections
Inside the front housing
1
1. Inactive switch
2. Active switch
3. Pressure connection
4. LCC-S connection
5. Output terminal
block
6. Power supply
terminal block
7. Cable gland
Removable front face
Fixed back housing
4
7
6
5
2
3