5
● Keep the batteries out of the reach of children. If a child accidentally swallows a battery, contact a
doctor immediately.
● Since the unit is designed to be sealed, rechargeable batteries must not be used. Gas could be
generated in the battery, resulting in an explosion or fire.
5-2. Installing a Vinyl Cover for the Unit
The SK-810PT comes with a vinyl cover to keep the unit clean and protect it from damage if dropped.
(1) Open the mouth of the cover and place the unit inside so that the sides of the vinyl cover having a slit
is over the back of the unit.
(2) Pass the connector on the top of the unit through the hole on the upper surface of the vinyl cover, and
then insert the strap at the end of the cover into the slit to close the cover.
Note: Static electricity could build up under the vinyl cover, resulting in unstable readings.
In such cases, remove the cover before using the unit.
5-3. Connecting the Sensor Probe
(1) While the unit is off, insert the sensor connector into
the unit by aligning the
凸
(concave) part of the
sensor connector with the
凹
(convex) part of the
connector on the unit.
(2) Turn the nut on the sensor connector clockwise to
secure the connector.
(3) To disconnect the sensor probe, turn the nut
counterclockwise and pull the sensor connector
straight out.
- To protect the sensor, make sure that the unit is off before installing or removing the sensor probe.
Caution
● Do not forcibly pull or turn the sensor connector.
● Do not remove the sensor probe by pulling the sensor cord.
5-4. Measuring Temperature
(1) Connect the sensor probe and turn on the unit.
- The "E_5" code appears if the unit is turned on without the sensor probe connected.
(2) Insert the sensor probe into the object to be measured.
- For a highly accurate temperature measurement, allow enough insertion depth so as not to be
affected by the temperature around the protection tube.
(3) Turn off the unit after use.
凸
(concave) part
凹
(convex) part