SATO 8312-00 Instruction Manual Download Page 2

 

 

Introduction 

 
Thank you for purchasing the Heat Stress Monitor Model SK-170GT. 
This product measures WBGT index*, air temperature, globe temperature, and humidity. Do 
not use it for other purposes. 
* The WBGT (Wet Bulb Globe Temperature) index is a general heat stress index used in ISO 

7243 

Hot environments - Estimation of the heat stress on working man, based on the WBGT 

index

 and its application in Japan, JIS Z8504. 

The WBGT index is often used in health and safety guidelines for workplaces and sporting 
situations. 

Please read this manual thoroughly before using the instrument and keep the manual in a 
safe place for your future reference. 

 

PRECAUTION

 

Make sure you fully understand the following facts about this instrument before use for proper 
usage. 

 

This instrument 

does not

 prevent heat stress. 

 

This instrument indicates WBGT index which can be used as guidelines for heat stress. 

 

Heat stress is caused not only by high  temperature ambient, but also by one’s health 
condition, one’s adaptability to hot environment, etc. 

 

WARNING 

Never use this instrument in an atmosphere exposed to inflammable gases as it 
may explode. 
* If there is anything you are not clear about this instrument, contact a dealer or 

us. 

 

CAUTION 

Please keep the followings for proper use of the instrument. 

 

Use this instrument in a normal atmosphere only. 

 

This instrument is not waterproof. Never let the instrument to get wet. 

 

Using the instrument out of its measuring range can cause malfunctions. 

 

Never directly touch the humidity sensor. 

 

If the instrument gets condensed with moisture, turn the power off immediately and let it 
dry naturally before using it again. 

 

If the instrument is used in an electrically noisy environment, the instrument could show 
unstable values and errors in accuracy. 

 

Never disassemble or alter the instrument to avoid any malfunctions and damages. 

 

This is a precision instrument. Be careful not to drop or give shocks to it. 

 

Summary of Contents for 8312-00

Page 1: ...INSTRUCTION MANUAL No 8312 00 Heat Stress Monitor Model SK 170GT JIS Conformed...

Page 2: ...idelines for heat stress Heat stress is caused not only by high temperature ambient but also by one s health condition one s adaptability to hot environment etc WARNING Never use this instrument in an...

Page 3: ...uideline for proper working environment Features JIS B7922 class 2 compliant Compliant with JIS B7922 class 2 electronic Wet Bulb Globe Temperature WBGT index meter It can be used as a guideline for p...

Page 4: ...the WBGT index reaches the limit set in advance Auto power off function Shuts down automatically in 20 minutes to save the battery Transmittable with RS232C Optional RS232C cable is available to send...

Page 5: ...mode press the key for 2 seconds to set the wind speed After changing them press the key again for 2 seconds to fix the value and return to measurement mode IN_OUT Next Press this key for 2 seconds in...

Page 6: ...TA globe temperature TG or humidity RH Also it displays the set value in alarm or wind speed setting mode MIN Indication for minimum measured value MAX Indication for maximum measured value WBGT Indi...

Page 7: ...ction by turning the cover Turn on the power by pressing the POWER key Select a measurement mode by pressing the MODE Up key The display changes in the following order WBGT index WBGT Air temperature...

Page 8: ...d speed to calculation For a better measurement set the ambient wind speed properly Note This instrument does not measure wind speed Press the POWER key to turn on Select the globe temperature TG mode...

Page 9: ...out 5 8 8 0 m s 13 18 mph Wind moves small branches Wind raises dust and loose paper from the ground and drives them along Abstracted from the reports from National Weather Service U S Dept of Commerc...

Page 10: ...r 20 minutes In case of using it for more than 20 minutes release the auto power off function Refer to Auto Power Off Function for details Note The buzzer does not work when MIN MAX values are display...

Page 11: ...oftware exclusively made for this product Provide software by yourself when needed How to Connect Plug the RS232C cable into the connection jack on the right side of the unit Connect the 9 pin connect...

Page 12: ...s missing Attach the globe to the main unit properly One of the measured values other than WBGT is over or under the display range Use within the measuring range E33 is displayed in the other measurem...

Page 13: ...e ambient before use Electric noises generated from the environment affects the state Do not use the unit in the circumstances where electric noises are generated A value around 0 0 C blinks in TG mod...

Page 14: ...pacitance sensor WBGT Alarm Setting Range 20 0 to 49 9 C 0 1 C step no deactivation Wind Speed Setting Range 0 1 to 3 0 m s 0 1 m s step Conformed Standard JIS B7922 Class 2 Approved measuring range w...

Page 15: ...ght prepaid basis and correction of the defect will be made without charges We alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repa...

Page 16: ...3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan https www sksato co jp en...

Reviews: