6
* If the SD character is flashing, it indicates that there is not enough free space. (It flashes when the free
space of the SD card falls below approximately 10MB.)
If the FULL character is lit, it indicates that there is no free space. Free up some space and reinstall the
SD card.
●
Removing the SD card
1) Turn the power off.
2) Remove the lid of the SD card slot from the back of the case by sliding it.
3) Push the SD card to the far end until a "click" is heard, then the spring fitted inside pushes out the card.
4) Remove the SD card by pulling it straight out.
5) Slide the lid into the unit until a "click" is heard to secure it.
●
About the SD card
In addition to the SD card provided, you can use a commercial SD or SDHC card of 1 to 8 GB in the unit.
Use a formatted SD card.
* NTFS cannot be used in the unit.
* We have verified the operation of the following SD cards:
SanDisk 2GB
"SDSDB-2048-J95B"
4GB
"SDSDB-4096-J95A" (comes with the product, formatted)
Panasonic 2GB "RP-SDL02GJ1K"
Transcend 2GB "TS2GSDC"
Attention
!
- We recommend using the provided SD card or one of the types verified by us. Using an SD card other than
these could result in a recognition error.
- Do not remove the label stuck to the SD card, nor attach any labels or stickers to it.
Before connecting or disconnecting the SD card, be sure to turn off the power of the unit.
Connecting or disconnecting the SD card while the power is on could result in failure of the unit or the SD
card, as well as loss of or damage to the data.
- We recommend backing up any data on the SD card to a device such as a PC.
Note that we are not responsible for any loss of or damage to data on the SD card.
Installing the probe
The probes with K-type thermocouple sensor can be connected to the unit.
If you use probes other than those recommended by us, be sure to use the ASTM (former ANSI) miniature
connector for the K-type thermocouple.
Note: The unit is designed to be used with the K-type thermocouple. Temperatures cannot be measured
properly if a probe other than the K-type thermocouple or a connecter other than for the K-type
thermocouple is used.
Summary of Contents for 8191-00