11
MEMO function
This function is used to make a memo of the current measured value.
There are two types when making a memo: one-time data memo and continuous data memo.
[Memo operation]
・
The measured values are saved on the SD card as a text format file.
When the MEMO function is used for the first time, the file [SK001.TXT] is created. Every time a memo is
made thereafter, data is added to the next line.
Note: The file saved by the MEMO function can be analyzed using a commercial spreadsheet program.
However, the software program "4ch DATALOGGER for Windows" that comes with the unit
cannot be used for analyzing the file.
Data Format
MN
: MEMO function
Interval : Blank space
Unit
: C (°C)
・
The MEMO function cannot be used when the
SD
character is off or when the
FULL
character or the low
battery indicator is lit.
・
If the [MEMO] key is pressed when any of the functions MAX/MIN, HOLD, T1 – T2, or REL has been
activated, logging is performed upon cancelling each function.
* If the offset value has been set, the values reflecting the offset value are logged.
●
One-time data memo
The measured value at the time when the [MEMO] key is pressed is logged on the SD card.
1) Press the [MEMO] key in the measurement mode.
The
character stays lit for five seconds to indicate that the measured value is being logged.
2) Upon completion of logging of the measured value, the measurement mode is resumed.
●
Continuous data memo
The measured values are logged as long as the [MEMO] key is kept pressed.
In the continuous data memo, the measured values are logged at two-second intervals.
1) Press and hold the [MEMO] key in the measurement mode.
The
character lights up to indicate that the measured values are being logged.
MN/AT date time int 1ch 2ch 3ch 4ch unit
MN 2011/1/2 12:55:21
155.5 300.5 658.4 1357 C
MN 2011/1/2 13:55:54
155.1 300.1 653.2 1341 C
MN 2011/1/4 14:54:48
154.9 299.5 640.2 1256 C