7-4. Setting the upper and lower thresholds for warning
This function is used to turn on the Alarm mark (Hi or Lo) when the
temperature or humidity threshold is triggered.
Note: No sound is produced while the Alarm mark is lit.
Tip: Aside from the steps below, you can set this function from
DATALOGGER for Windows. For further information, please refer to the
online Help.
(1) Press and hold the POWER ON/OFF key for one second or more to
turn on the Datalogger.
The LCD turns on to indicate that the SK-L200T II or SK-L200TH II
α
is
in the measurement state for temperature or temperature and humidity,
respectively.
Set the upper limit for the temperature first.
(2) Press the MODE key nine times.
(3) The Alarm and Hi marks light. Press the
▲
and
▼
keys at the same time.
Tip: Pressing the
▲
key and
▼
key simultaneously one more time will turn
off the warning function.
(4) Use the
▲
or
▼
key to display the desired value for upper limit.