−43 −
・
The logging option was “Ends at...” (limited amount) and logging ended
because the quantity of logging data reached the set end-point.
・
The logging option was “By pages” (partitioning) and the logging
ended because the memory became full (2000 readings in each page
of four pages).
To download data, turn on the Datalogger again.
●
Download all the necessary data at a time and do not leave any out.
Data stored on the Datalogger will be erased once logging starts again.
●
When setting the parameters using the software or while downloading is
in progress, do not disconnect the cable or turn the power off and then
on again, as this may cause a setting error and/or missing logged data.
●
Before erasing the recorded data, make sure that the low battery mark is
turned off. Do not remove the batteries while erasing data, otherwise the
erasing may fail.
If the battery power runs out while data is being erased, immediately
replace the batteries with new ones, and then follow the steps below to
resume operation.
①
Turn on the unit and press and hold the REC key for at least two
seconds until the REC mark on the display lights up.
②
Press and hold the REC key again for at least two seconds to stop
recording.
③
Erase the recorded data (refer to step (9) of page 42).
The memory bars on the display turn off after blinking to indicate that the
data has been successfully erased.
Summary of Contents for 8161-00
Page 33: ...31 6 Confirm Installation window appears Click next to start the installation...
Page 51: ...49 4 From the Setting menu click ID D 5 The confirmation message appears Click OK...
Page 68: ...66 Create and type an easily identifiable name for the data...
Page 79: ...77 ACCURACY OF R HUMIDITY...