4
How to Use
●
Installation location
Install the instrument indoors. It is available to mount it on a wall or a celling (option)
Avoid following locations for installation:
- Places with flammable gas, explosive gas and corrosive gas such as SO
2
and H
2
S
- Places where the ambient temperature go beyond 0 to 50
℃
(humidity of less than 85%rh)
- Places of the impact of electrical noise and radio waves
- Places in effects with condensing, rain, water drops or steam
●
Installation
Four wall-mounting brackets are included with the product.
If the product is to be suspended, use the two ceiling-mounting brackets (options). To purchase the
brackets for celling-mounting, contact the store where you purchased the unit or us.
1. Wall mounting
(1) Remove the eight M3 screws from the back of the unit case.
- Do not loosen or retighten the screws other than the eight shown in the figure below.
Back view
(2) Align the screw holes on the four wall-mounting brackets included with the product with the screw
holes on the unit, and then secure the brackets with the eight screws removed in step (1) above.
- Be sure to use the M3 screws removed from the unit.
Back view