5
2. Wiring method
(1) Loosen the six screws at the rear side of the instrument and remove the cover.
* Do not loosen or re-tighten the screws other than those six screws.
(2) Insert the cable through from the outside of the cable wiring port on the cover and wire to
the terminal block as shown in the drawing below.
* Cut an opening to the film covering the cable wire port
with a cutting knife.
* Please note the direction of the cover
○ Wiring diagram
<with the cover>
<without the cover>
Cover fixture screws (size: M3)
Cover fixture screws (size: M3)
terminal block
○ Terminal description
Terminal
name
Description
T+
Temperature current input + Terminal
T-
Temperature current input - Terminal
H+
Humidity current input + Terminal
H-
Humidity current input - Terminal
Both ends are for terminal block fixture screws.
Do not loosen the screws.
Temperature
Humidity