4
How to use
1. Installation
SK-500B is an indoor use desk-top digital barometer. Install the equipment horizontally avoiding the
places as below.
In flammable gas, explosive gas and corrosive gas
A place subject to direct sunlight and near heaters
At the exhaust port of air conditioner
In a place subject to vibration
In outside of the operation ambient (-10 to 50°C, less than 85%rh, no condensing)
In an envelopment with electric noise
Near splashing water and the place rainwater pours.
2. Connection
2-1. Pressure port
This is the inlet for atmospheric pressure. An R1/4 fitting can be used for connection. When
connecting a fitting, use a sealing material such as sealing tape to prevent leakage.
- Fitting tightening torque: 10 to 12 Nꞏm –
Caution
• Do not allow any foreign matter such as the end of the sealing material to enter inside the pressure
port. If foreign matter gets inside, measurement errors or failure could result.
• When installing the fitting, do not overtighten it as this could result in failure.
Be sure to tighten it within the tightening torque range.
2-2. RS-232C/Printer output
When the RS-232C/Printer output is to be used, first connect the cable below to the RS-
232C/Printer output port on the back of the unit.
When a USB-Serial cable is used for the first time, you may have to install the USB driver.
For details on installing the USB driver and its system requirements, refer to the instruction manual
of the USB-Serial cable that you will use.
Communications standard for RS232C and Printer output
Connection
Connection cable / connector
Personal
computer
Serial connecting terminal
(D-sub 9 pin, male)
Attached RS232C cable (D-sub 9 pin, crossing)
USB connecting terminal
(Type-A, female)
Attached RS232C cable (D-sub 9 pin, crossing)
and USB serial cable
Exclusive printer Serial connecting terminal
(D-sub 9 pin, female)
Attached cable for printer (D-sub 9 pin, straight)
Note: We have checked the operation with the USB-Serial cables (Japanese version) listed below.
Depending on the use of other USB-serial cable, the country to be used and the connection
environment, the unit may fail to operate correctly. Our technical and customer supports are
provided to Japanese OS only.
BSUSRC0605BS (cable length: 0.5 m) of Buffalo Inc. (supplied with the product)
BSUSRC0610BS (cable length: 1.0 m) of Buffalo Inc.