Heat-running and zero point adjustment 

(1)  Place the Function Switch to either of Air Velocity or Static Pressure position. Then put the 

sensor cap on the probe and run the heat for more than 5 minutes. The pointer will gradually 

return to zero point from left (minus). 

This is the time required to stabilize after electric current flows through the circuit and the 

sensor.     

(2)  If the pointer does not point to zero of the scale after the heat-running, set the pointer to zero 

by turning the rotary switch of zero point adjustment. 

For fine adjustment, place the Range Selector Switch to HIGH first. Then switch the Range 

Selector Switch to LOW position and adjust it again.   

Note

: Be sure to make the zero point adjustment under normal temperature (at 20°C±15°C). 

If the pointer fully points to the left (minus) or right (plus) end, turn function switch to 

Power-Off position and contact us since it will not be normal.   

 

 

Measurement range (Range Selector) 

Measuring ranges of air velocity, temperature and static pressure are as follows.   

Select HIGH and LOW depending on the circumstances. 

Range 

  LOW 

HIGH 

Air velocity 

0.5 to 3 m/sec. 

3 to 40 m/sec. 

Temperature 

0 to 60°C 

60 to 120°C 

Static pressure 

0 to 500 Pa 

500 to 4500 Pa 

 

 

How to measure air velocity 

(1)  Place Function Switch in Air Velocity position. Put the sensor cap on the probe and make 

zero point adjustment.

 

Note: 

Place Range Selector Switch in HIGH position when air velocity cannot be estimated. 

 

(2)  Measure air velocity facing the wind direction mark of the probe 

(red arrow mark under the detection window) to windward.

 

Note: 

Face the detection window vertically against the wind 

direction. If the mark is aslant against the wind direction, the 

air velocity will be indicated higher than the actual value. 

 

* When the ambient temperature to be measured is high, it will take some time until the 

temperature of sensor section rises to the ambient temperature (about 3 minutes at 100°C). 

Take measurement after the temperature of sensor section is fully stabilized. Also, when 

taking measurement in a normal ambient temperature in LOW range (lower than 3 m/sec.) 

wait until the temperature is fully stabilized. 

wind 
 
 
 
wind   
direction   
mark 

Summary of Contents for SK-28V

Page 1: ...Instruction Manual No 7683 01 Hot Wire Anemometer Model SK 28V L10...

Page 2: ...the SK 28V please contact our retailer or our service network Descriptions Adopting the principle of thermal anemometer this product can measure wide range of air velocity The measured value can be r...

Page 3: ...r Lamp Function Switch Range Selector Zero Adjustment Pivot Pointer Sensor Cap Detection Window with mesh Air velocity Sensor Temperature Sensor Wind Direction Mark Sensor Cord Carrying Case Sensor St...

Page 4: ...pacity 1 Place the Function Switch in BATT position Confirm that the pointer is in the blue zone of the scale plate When the batteries are new the pointer will point to the right end plus side which i...

Page 5: ...ature and static pressure are as follows Select HIGH and LOW depending on the circumstances Range LOW HIGH Air velocity 0 5 to 3 m sec 3 to 40 m sec Temperature 0 to 60 C 60 to 120 C Static pressure 0...

Page 6: ...ng Make sure to face the wind direction mark to the sucker for negative pressure Refer to Fig 1 4 Strongly press the sucker for negative pressure against the measurement hole and take measurement Refe...

Page 7: ...ccuracy of measured value of air velocity will be affected In such a case contact us for cleaning unit on charge bases 4 Never wipe or rub the unit with a dry cloth or like for cleaning Doing so may g...

Page 8: ...ement 9 0VDC C size alkali battery LR14 6 pcs Battery Life About 20 hours in continuous use at air velocity of 20m s at 25 C in use of new C size batteries Sensor Element Specially coated platinum ele...

Reviews: