9
3. Clearing data in memory
(1) Turn off the SK-1110.
(2) Press the MEM and PWR keys together to turn on the unit. Release the MEM key
when the display shows the following image.
(3) Use the
▼
key to select a memory clearing option.
Choose YES or NO in the lower display.
"YES" ------ Clear memory and resume measurement.
"NO" ------- Resume measurement without clearing memory.
(4) To resume measurement, press the MEM key.
NB
The data stored in memory will not be lost even the unit is turned off or the battery is
removed.
Error Message
Display Causes
Remedy
“--------“
Probe is not connected
Connect the probe. K type thermocouple is
applicable
Incorrect thermocouple sensor is
connected
Connect the K type thermocouple sensor
OL or -OL
Measured value is lower or more
than the display range.
The display range of the unit is from -73.3°C
to 1371.0°C. Bring the value within the
measuring range