page 5
Manual SA 235
Switching on
Switch the scale on by pressing key for at least 1 second.
The weighing platform has to be empty !
A "beep" sounds and the display shows all the Led's.
During the start up procedure the software version number is shown.
After that the date of the software appears.
Now the scale shall automatically set zero, as soon as there is an exact
standstill of the weight signal.
The Led's at : and at the weight unit light up.
If the weight signal is not stable, or a defect in the scale is detected,
the display stays at - - - - - - .
After maximal 10 minutes warming up time the system is stabilized and the
scale is ready for use.
If the scale with empty platform is not at zero, press key .
The zero setting range is + and - 2% of the scale capacity.
Outside this range a zero setting shall not be executed
At bigger deviation from zero: switch the scale off and on again.
There is no limitation for the zero setting at switching on.
Please attention ! The available measuring range keeps the same: so the
loadcell may be destroyed if the zero setting value is too high !
Place empty container on the scale and tare to zero with key
If the weight reading is not stable the tare is not executed !
The Led above indicates the taring: reading shows the net weight.
This may be repeated up to the maximum weighing capacity.
Delete the tare with an empty weighplatform with key .
Place the object with the correct weight on the scale, press key .
The reading shows 0 . The Led above: lights up.
Place the object to be checked on the scale and read the deviation in + or - .
Delete the tare with an empty weigh platform with key .
WEEGTECHNIEK HOLLAND B.V.
Website www.satex.nl
Tel +31 36 522 20 30
Patroonsweg 23
Email [email protected]
Fax +31 36 522 20 60
3892 DA ZEEWOLDE The Netherlands
Set zero
Tare
(taring is setting zero within the total weighing range)
Plus / minus checkweighing
lb
kg
g
lb
kg
g
lb
kg
g
lb
kg
g
lb
kg
g
lb
kg
g
lb
kg
g
lb
kg
g