Manual  SA 235 
 

Switching on   

 
 
 
 
 
 
 
 
 
 
 

 
Switch the scale on by pressing key          for at least 1 second. 
The weighing platform has to be empty ! 
A "beep" sounds and the display shows all the Led's.  

 
 

 

 
 

 

 

 

 

 

 

During the start up procedure the software version number is shown. 

 
 
 

  

After that the date of the software appears. 
 
 
Now the scale shall automatically set zero, as soon as there is an exact 
standstill of the weight signal.  
The Led's at :           and at the weight unit light up. 
 
If the weight signal is not stable, or a defect in the scale is detected, 
the display stays at - - - - - -  .  
After maximal 10 minutes warming up time the system is stabilized and the 
scale is ready for use. 
 
 
 
 
 
If the scale with empty platform is not at zero, press key          . 
The zero setting range is + and - 2% of the scale capacity.  
Outside this range a zero setting shall not be executed 
 
At bigger deviation from zero: switch the scale off and on again.  
There is no limitation for the zero setting at switching on.  
Please attention !  The available measuring range keeps the same: so the 
loadcell may be destroyed if the zero setting value is too high ! 
 
 
 
 
Place empty container on the scale and tare to zero with key  

 

If the weight reading is not stable the tare is not executed ! 

The Led above        indicates the taring: reading shows the net weight

    

This may be repeated up to the maximum weighing capacity. 

Delete the tare with an empty weighplatform with key          . 

  

 
 
 
Place the object with the correct weight on the scale, press key           .  

The reading shows 0 . The Led above:       lights up. 

 

Place the object to be checked on the scale and read the deviation in + or - . 

Delete the tare with an empty weigh platform with key          . 
 
          

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WEEGTECHNIEK HOLLAND B.V. 

 

 

Website   www.satex.nl   

 

 

 

Tel    +31 36 522 20 30

 

Patroonsweg 23 

 

 

 

 

Email       [email protected]  

 

 

 

 Fax  +31 36 522 20 60 

3892 DA   ZEEWOLDE  The Netherlands 

Set zero  

Tare

                                        (taring is setting zero within the total weighing range) 

Plus / minus checkweighing

              

lb 

kg 

lb 

kg 

lb 

kg 

lb 

kg 

lb 

kg 

lb 

kg 

lb 

kg 

lb 

kg 

Summary of Contents for SA 235

Page 1: ...l and high resolution of the reading user selectable Weegtechniek Weegtechniek Weegtechniek Weegtechniek holland b v holland b v holland b v holland b v Patroonsweg 23 3892 DA Zeewolde The Netherlands Tel 31 36 522 20 30 Fax 31 36 522 20 60 Email info satex nl Web www satex nl ...

Page 2: ... responsibility if the guidelines of this manual are not observed Before any work is done on the scale power connections from the mains adaptor and the internal battery should be interrupted Warranty is void if the scale is opened anyhow The instrument fulfils the requirements laid down in the EG regulations 89 336 EWG and 73 23 EWG for electromagnetic compatibility Exceeding of the maximum values...

Page 3: ...s and scale divisions resolution carrying grip cable locker with slide door bottom side Reading 6 decades Led weight display figures 15 mm high 3 check weigh Led s 3 battery status Led s 6 functional Led s Display speed speed and filter functions separately adjustable Scale divisions selections standard H1 2 x more accurate H2 5 x more accurate H5 Housing Stainless steel housing and weighplatform ...

Page 4: ...her the weighing range battery is sufficiently loaded near empty Lo of empty Cursor increase figure with 1 At Lo the scale switches off after 15 minutes In the User Mode the accuracy of the reading may be selected All versions have three selectable multi interval readings with the weighing capacity divided into two ranges Example for a 7 5 kg scale see table on page 3 Reading type H1 1 st range 0 ...

Page 5: ... and on again There is no limitation for the zero setting at switching on Please attention The available measuring range keeps the same so the loadcell may be destroyed if the zero setting value is too high Place empty container on the scale and tare to zero with key If the weight reading is not stable the tare is not executed The Led above indicates the taring reading shows the net weight This ma...

Page 6: ...ss key 2 seconds until the display shows rANG Now release both the keys 2 Press again key in order to enter the Low weightvalue margin the Led next to LO lights up 2 The flashing figure may be increased with cursor key With cursor key the next to the right figure can be selected 3 Confirm the entered weightvalue with enter key The display now switches to input of the High weightvalue margin the Le...

Page 7: ...after start up of the program 7 The scale is now ready for piece counting The Led above the symbol is on counting mode The article weight is memorised it will not be deleted if the scale is switched off 8 Key switches off temporarily and on again the counting mode Note if the reading is in the unit If temporarily switched off the Led above keeps on indicating that gram it will be switched automati...

Page 8: ...isions Improved accuracy with 2 x 6000 scale divisions e g 0 1 5 kg with 0 2 gram and 1 5 3 kg with 0 5 gram divisions Confirm the selection with key the next menu section will be switched on In the multi interval weighing modes after taring with key the higher resolution of the 1 st range is started again even in the 2 nd weighing range Selectable with key are 3 units of display weight reading 0 ...

Page 9: ...tes of standstill not in the user and calibration modes 0 auto off is switched off 1 2 minutes 2 5 minutes 3 20 minutes 4 60 minutes The delay time may be selected with key Confirm the selection with key the next menu section will be switched on Connected with the signs HI OK and LO in the check weighmode a interrupted beep may sound 3 settings are available 0 no beep at all 1 beep only if the wei...

Page 10: ...ion with key hereafter the user menu is closed After changing one of the settings in the user mode all menu functions have to be passed through with enter key in order to close the menu and enter the settings that have been changed After the last menu function SPd the scale switches off and returns after some seconds in the weigh mode If the user menu is not closed in this way but in an earlier st...

Page 11: ...ver regard the capacity of the loadcell Never select a higher capacity it will cause destruction of the loadcell 7 confirm with key het display toont LS 8 start the zero point adjustment using key again 9 now the zero adjustment is executed the most right display figure runs down from 5 to 0 10 continue with the 1 point or the 3 points adjustment procedure depending on the display showing H2 or H1...

Page 12: ... with accurate weights Repeat the adjustment procedure if necessary The weigh platform is overloaded a warning beep sounds Remove the overload as soon as possible Weight reading is gross negative Zero setting and taring is not possible The scale has been switched on with a load on the platform and has set zero with this load After that the load has been removed and now the weight reading is negati...

Reviews: