Satex SA 235 User Manual Download Page 2

page 2 

 

Manual  SA 235 

 

 

 

 

 
 
 
 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

         

 

PCG/PC SA 235 Manual 14-03-2012 

 

General 

 

Thank you for purchasing this waterproof SATEX precision weighing scale ! 
This scale is equipped with reliable technology and has user friendly operational software. 
You will use it many times and it will prove itself to be an easy, very speedy and accurate tool. 
 
If any questions raise that are not described in this manual, please address yourselves to your SATEX supplier, 
have a look at our website : www.satex.nl  or send your questions per email to : [email protected] 

 
Application,  waterproof,  HACCP
 

 
The water resistant SA 235 precision scale is designed for :  

 

 

-  use in wet and damp condtions 

 

 

 

-  very accurate (high resolution reading) and very fast weighing 

 

 

-  check weighing with 2 settable weight margins 

 

 

-  counting pieces  

 
The scale may be used without problems in very humid and moist conditions. 
The scale fulfils the requirements of Norm EC 60529:2001, sealed to IP 67: submerged in clean water during 
1 hour at a depth of 1 m, penetrating water will not cause malfunctions. 
 
The stainless steel exterior of the scale is designed according to the HACCP principles for food safety.  
 
The  SA 235 is designed for internal weighing only. The use of the scale is not permitted for commercial purposes. 
 

Safety

 

 
Please read this manual thoroughly before start using the scale. 
The supplier does not accept any responsibility if the guidelines of this manual are not observed. 
 
Before any work is done on the scale, power connections from the mains adaptor and the internal battery 
should be interrupted. Warranty is void if the scale is opened anyhow.  
 
The instrument fulfils the requirements laid down in the EG regulations 89/336/EWG and 73/23/EWG for 
electromagnetic compatibility.  
Exceeding of the maximum values, laid down in these regulations, must be avoided.  
The instrument is isolated according to class 2:   
  
Electrostatic discharges may cause fatal damage to the scale and should be avoided at all times.  
A.o. a proper earth connection of the scale and the place of erection is required in such cases. 
 
In case the 230 V excitation cable is damaged, or the scale is not working with the 230 V connected, the 
scale should not be used any longer and sent for service to your supplier. 
 
The built in sealed lead battery should, if it has to be replaced, be disposed off as small chemical waste,  
not to be added to normal waste.  
 

Maintenance

 

 
Except from regular cleaning with a slightly humid cloth and a not aggressive detergent, further maintenance 
is not required.  
However it makes sense to check the scale’s weighing accuracy with regular intervals, using accurate calibration 
weights of class M1 at least. The scale may be adjusted by the user itself using accurate class M1 weights. 
(see: Adjusting the scale) 
 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WEEGTECHNIEK HOLLAND B.V. 

 

 

Website   www.satex.nl   

 

 

 

Tel    +31 36 522 20 30

 

Patroonsweg 23 

 

 

 

 

Email       [email protected]  

 

 

 

 Fax  +31 36 522 20 60 

3892 DA   ZEEWOLDE  The Netherlands 

Summary of Contents for SA 235

Page 1: ...l and high resolution of the reading user selectable Weegtechniek Weegtechniek Weegtechniek Weegtechniek holland b v holland b v holland b v holland b v Patroonsweg 23 3892 DA Zeewolde The Netherlands Tel 31 36 522 20 30 Fax 31 36 522 20 60 Email info satex nl Web www satex nl ...

Page 2: ... responsibility if the guidelines of this manual are not observed Before any work is done on the scale power connections from the mains adaptor and the internal battery should be interrupted Warranty is void if the scale is opened anyhow The instrument fulfils the requirements laid down in the EG regulations 89 336 EWG and 73 23 EWG for electromagnetic compatibility Exceeding of the maximum values...

Page 3: ...s and scale divisions resolution carrying grip cable locker with slide door bottom side Reading 6 decades Led weight display figures 15 mm high 3 check weigh Led s 3 battery status Led s 6 functional Led s Display speed speed and filter functions separately adjustable Scale divisions selections standard H1 2 x more accurate H2 5 x more accurate H5 Housing Stainless steel housing and weighplatform ...

Page 4: ...her the weighing range battery is sufficiently loaded near empty Lo of empty Cursor increase figure with 1 At Lo the scale switches off after 15 minutes In the User Mode the accuracy of the reading may be selected All versions have three selectable multi interval readings with the weighing capacity divided into two ranges Example for a 7 5 kg scale see table on page 3 Reading type H1 1 st range 0 ...

Page 5: ... and on again There is no limitation for the zero setting at switching on Please attention The available measuring range keeps the same so the loadcell may be destroyed if the zero setting value is too high Place empty container on the scale and tare to zero with key If the weight reading is not stable the tare is not executed The Led above indicates the taring reading shows the net weight This ma...

Page 6: ...ss key 2 seconds until the display shows rANG Now release both the keys 2 Press again key in order to enter the Low weightvalue margin the Led next to LO lights up 2 The flashing figure may be increased with cursor key With cursor key the next to the right figure can be selected 3 Confirm the entered weightvalue with enter key The display now switches to input of the High weightvalue margin the Le...

Page 7: ...after start up of the program 7 The scale is now ready for piece counting The Led above the symbol is on counting mode The article weight is memorised it will not be deleted if the scale is switched off 8 Key switches off temporarily and on again the counting mode Note if the reading is in the unit If temporarily switched off the Led above keeps on indicating that gram it will be switched automati...

Page 8: ...isions Improved accuracy with 2 x 6000 scale divisions e g 0 1 5 kg with 0 2 gram and 1 5 3 kg with 0 5 gram divisions Confirm the selection with key the next menu section will be switched on In the multi interval weighing modes after taring with key the higher resolution of the 1 st range is started again even in the 2 nd weighing range Selectable with key are 3 units of display weight reading 0 ...

Page 9: ...tes of standstill not in the user and calibration modes 0 auto off is switched off 1 2 minutes 2 5 minutes 3 20 minutes 4 60 minutes The delay time may be selected with key Confirm the selection with key the next menu section will be switched on Connected with the signs HI OK and LO in the check weighmode a interrupted beep may sound 3 settings are available 0 no beep at all 1 beep only if the wei...

Page 10: ...ion with key hereafter the user menu is closed After changing one of the settings in the user mode all menu functions have to be passed through with enter key in order to close the menu and enter the settings that have been changed After the last menu function SPd the scale switches off and returns after some seconds in the weigh mode If the user menu is not closed in this way but in an earlier st...

Page 11: ...ver regard the capacity of the loadcell Never select a higher capacity it will cause destruction of the loadcell 7 confirm with key het display toont LS 8 start the zero point adjustment using key again 9 now the zero adjustment is executed the most right display figure runs down from 5 to 0 10 continue with the 1 point or the 3 points adjustment procedure depending on the display showing H2 or H1...

Page 12: ... with accurate weights Repeat the adjustment procedure if necessary The weigh platform is overloaded a warning beep sounds Remove the overload as soon as possible Weight reading is gross negative Zero setting and taring is not possible The scale has been switched on with a load on the platform and has set zero with this load After that the load has been removed and now the weight reading is negati...