11
F
AULT
No operation
No spray
No water at tip but
vibration present
Insufficient power
Low vibration
No ultrasonic output
Water leak between
handpiece and cord or at
front/back junction of
handpiece
R
EMEDY
• Check mains outlet
• Return to customer
service
• Dry any moisture, especially
on electrical contacts
• Check water supply
• Check pressure
• Clean or replace filter
• Return to customer
service
• Remove obstruction
• Adjust spray
• Replace tip
• See tip utilization method
• Wipe electrical contacts
dry
• Tighten with wrench
• Clean cord contacts
• Return to customer
service for cord
replacement
• Replace seal
P
OSSIBLE CAUSE
• Wrong mains cord connection
• Blown fuse
• Presence of fluid or moisture between
handpiece and cord
• Wrong water tube connection
• No water pressure
• Clogged filter or defective solenoid
valve
• Blocked tip
• Incorrect spray adjustment
• Worn or deformed tip
• Incorrect use: wrong angle of
application or inadequate pressure on
tooth
• Presence of fluid or moisture between
handpiece and cord
• Incorrect tip tightening
• Defective connector contact
• Cut wire in handpiece cord
• Worn sealing gasket (1.15 x 1) in
handpiece