Initial Operation of Balances Without a Draft Shield
Touch CAL to select
the combined
calibration and adjust-
ment mode.
Quick Access to
Calibration and
Adjustment
You will need to perform the combined calibration
and adjustment procedure the first time you operate
the balance, each time after changing it to a different
location and at regular intervals during operation.
If you are using a Practum series balance, you will
have to do this manually as described below.
The Secura and Quintix series perform this procedure
fully automatically.
To perform external
adjustment, touch
CAL-Extern.
First touch
A
, then
w
to calibrate
the balance. Place a
calibration weight
equal to that displayed
on the balance on the
pan. The balance will
now perform auto-
matic calibration and
adjustment, then it will
change to the operat-
ing mode.
Getting Started Quickly
Initial Operation of Balances with a Draft Shield
Install the following parts in
the order listed:
– In-use dust cover
– Turn the draft shield lock
to “unlocked” .
– Place the draft shield (1) on
the balance. Note: Use both
hands to hold the top struts
of the draft shield.
– Slide the side doors all the
way to the back.
– Turn the draft shield lock
to “locked” .
– Position the base plate (2),
centering ring* (3), pan
support (4), weighing pan (5)
and high-rim shield ring* (6)
on the balance.
Note: For information about
how to clean the draft shield,
see the Care and Maintenance
chapter in the Installation
Instructions.
* only for Secura with 0.1 mg
readability
Install the following parts in
the order listed:
1) In-use dust cover
2) Position the shield
disk (1) and turn it clock-
wise until it is locked
securely in place.
3) Position the pan support (2)
and the weighing pan (3).
For detailed instructions, please refer to our Installation
Instructions and our detailed User Manual.
Secura Balances
Quintix and Practum Balances