SARTORIUS MC 5 Balance
Stand 1/94
MC 5
Power-On Routine and Error Messages
Once your have turned on the balance using the
key, " 0.000 mg “ is displayed. This indicates that
the balance is ready to operate.
Error Messages
On the MC 5, error messages are indicated by a code number.
All codes for operating errors, including errors in the sequence of a function, have two-digit numbers.
Error codes for hardware errors have three-digit numbers.
The error codes enable you to identify the type of operating error.
Display
Meaning
Reason/Solution
Err 01
Display format overflow
The weighing system is overloaded.
Err 02
Zero point error at the start
Re-tare;
of the CAL routine
check system.
Err 04
Control error during
Check system for stability.
multiple calibration mode
Err 06
Internal calibration weight
Check system. Return the microbalance
defective or not available
to the factory, if necessary.
Err 10
1st tare memory disabled
The tare functions are
when a value is stored in
interlocked. Clear tare
2nd tare memory
memory to enable the CF/ tare key.
Err 11
Attempt to store an invalid
The readout was negative,
number in tare memoryfor example -
check sample.
using TAR -F1/
Err 20
The ></ key was pressed
For instance, you forgot
before an application
to store a reference
program was initialized
percentage for weighing in percent.
Err 21
Reference percentage or
Code 2 3 1 is selected in the balance
reference piece count setting
operating menu. Change this code
disabled, or this application
to 2 3 2 or 2 3 3.
does not have a parameter mode
Err 22
Reference cannot be stored
The weight is too light or
for weighing in percent or counting
no sample is on the pan.
Err 23
Value cannot be stored
No sample is on the pan or
for any particular application program
the readout is negative.
Err 30
The print key was pressed
Connect printer.
in the BPI mode; no printer on-line
Err 50
Converter error (TC range)
Check system or return microbalance to the
factory in Goettingen.
Err 53
Temperature sensor circuit failure
Check temperaturecompensating sensor or
return microbalance to the factory in Goettingen.
Err 54
Balance converter value below
Check system.
the minimum limit
Beam at lower stop.
Summary of Contents for MC 5
Page 1: ...Sartorius Balance Service Manual MC 5...
Page 4: ...4 SARTORIUS MC 5 Balance Stand 1 94 MC 5 Exploded View Diagram 1...
Page 6: ...6 SARTORIUS MC 5 Balance Stand 1 94 MC 5 Exploded View Diagram 2...
Page 8: ...8 SARTORIUS MC 5 Balance Stand 1 94 MC 5 Exploded View Diagram 3...
Page 10: ...10 SARTORIUS MC 5 Balance Stand 1 94 MC 5 Exploded View Diagram 4...
Page 12: ...12 SARTORIUS MC 5 Balance Stand 1 94 MC 5 Exploded View Diagram 5...