24
Specifications
Model
B 3100S
B 4100
B 6100
Capacity/weighing range
g
3,100
4,100
6,100
Readability g
0.01
0.1
0.1
Tare range (by subtraction)
g
-3,100
-4,100
-6,100
Standard deviation
g
≤
±0.01
≤
±0.05
≤
±0.05
Max. linearity
g
≤
±0.02
≤
±0.1
≤
±0.1
Stabilization time (typical)
s
2
1.5
1.5
Display update
s
0.1 – 0.8 (selectable)
Adaption to ambient conditions
and application requirements
by selection of one of 4 optimized filter levels
Stability range
d
0.25 ... 64 (selectable)
Ambient temperature range
K
273 – 313 (0
o
C ... +40
o
C)
Allowable relative humidity
15 … 85
(moisture-proof rating)
Sensitivity drift within
+10 … +30
o
C
/
o
C
≤
±2
∙
10
-6
≤
±5
∙
10
-6
≤
±5
∙
10
-6
Pan size
mm
in.
235 x 180
9.3 x 7.1
Balance housing (W x D x H)
mm
in.
250 x 270 x 65
9.8 x 10.6 x 2.6
Net weight
kg/lbs. 5/11
3.5/7.7
3.5/7.7
Power requirements
(v frequency)
115 or 230 V depending on the AC adapter used
50 – 60 Hz
Allowable voltage fluctuation
-20% … +15%
Power consumption(typical)
VA
max. 7.5
7.5
7.5
Interface (optional)
RS 232 C/V24 – 28, RS 423/V10; 7-bit;
parity: even, mark, odd, space;
transmissions rate: 150 ... 9,600 Baud