Setting IDs, Time, Date
and Display Brightness
Purpose
– Configuring measurement environment parameters
for ISO/GLP-compliant data records.
– Setting date and time (for ISO/GLP-compliant
records only).
– Adapting display to ambient lighting conditions.
Features
– Enter up to 8 characters to identify a measurement
series. Permissible characters include the numbers 0
through 9 and the dash or minus sign (“–"). A dash
is output as a space on printouts. Leading zeroes are
not output.
– Date and time at beginning and end of ISO/GLP-
printouts.
– Display brightness
1
):
0 = off; levels of brightness: 1 through 9
Key functions during configuration:
Activate input of IDs, time and date:
Switch the balance off and then on again by
pressing
e
; while all segments are displayed,
press the
v
key briefly
Scroll upward
↑
: Press
q
Scroll to the right
→
: Press
r
Confirm input and toggle between IDs, time and
date: Press
w
Save settings and exit menu:
Press and hold
w
(> 2 sec.)
1
) No display backlighting on models LE26P(-0CE)
and LE225D(-0CE)
30