Operating Instructions 
 
 
Controller for Automatic Mass Comparators 
from Software Version 1.22    
for 

CC.......- L / CCE.......- L Model 

 

 

 

 
 

 
 

 
 

 
 

 

 

 

 
 

 
 

 
 

 
 

 
 

 

98648-006-96 

Summary of Contents for CCEx-L series

Page 1: ...Operating Instructions Controller for Automatic Mass Comparators from Software Version 1 22 for CC L CCE L Model 98648 006 96...

Page 2: ...oward Mass Comparator 14 Move Horizontally 15 Starting the Mass Comparison 17 1 without start delay time is setting 17 Select the procedure for mass comparison 17 Begin Measurement 18 Continuous Funct...

Page 3: ...he same line to the function sequences on the touch screen Operators errors are virtually impossible The controller recognizes the position of the load alternator at all times if you select a function...

Page 4: ...es Vertical movement short press move from the top toward to centre or from the centre to the lower position or raise from the lower to the centre or raise from the centre to the upper position long p...

Page 5: ...ws model no of the connected comparator software version language setting date Press anywhere to access the main menu Initial System Start up load the weights Press anywhere to access the main menu C...

Page 6: ...sired language 1 Deutsch 2 English 3 Espa ol 4 Francais C C 1 0 0 0 0 U L S o f t w a r e V 1 22 E n g l i s h 1 J A N 2010 C C 1 0 0 0 0 U L S o f t w a r e V 1 22 E n g l i s h 1 J A N 2010 The time...

Page 7: ...ime 1 Language 2 Time for averaging 10 3 Date and time 4 More Configurations C C 1 0 0 0 0 U L S o f t w a r e V 1 22 E n g l i s h 1 J A N 2010 The procedure for setting the date and time is self exp...

Page 8: ...ent sequence The factory setting is 03 x A B B A To select a different number of ABBA cycles Press the numeric key 1 1 1 Set number of A B B A cycles 03 2 Set number of Pre Cycles 00 3 Delay time befo...

Page 9: ...the numeric key 2 2 C C 1 0 0 0 0 U L S o f t w a r e V 1 22 E n g l i s h 1 J A N 2010 1 Set number of A B B A cycles 03 2 Set number of Pre Cycles 00 3 Delay time before taken values 20 4 More Confi...

Page 10: ...asurement The factory setting is 20 To select a different number of Delay time Press the numeric key 3 3 C C 1 0 0 0 0 U L S o f t w a r e V 1 22 E n g l i s h 1 J A N 2010 1 Set number of A B B A cyc...

Page 11: ...visa versa Scale Timeout ON waiting time 1 hour Scale Timeout OFF waiting time infinity To change between O ON O OFF press the numeric key 1 1 Note After time out the display will show this message No...

Page 12: ...s C C 1 0 0 0 0 U L S o f t w a r e V 1 22 E n g l i s h 1 J A N 2010 Set a number of time between 000 and 255 Minute Then press to confirm new settings or to cancel new settings The display returns t...

Page 13: ...through several position shifts until it recognizes the initial raised position Always activate this function after turning on the mass comparator after canceling a measurement by pressing reset and...

Page 14: ...ight is placed on the weighing pan Press the number key 3 3 You can select this function immediately following Function 1 if desired 1 Raise load alternator to upper position 2 Move load alternator to...

Page 15: ...ty adjustment of the mass comparator return the load alternator platform to the upper position Press the number key 1 1 1 Mass Comparison cycles 2 Move vertically 3 Move horizontally 4 Configure C C 1...

Page 16: ...U L S o f t w a r e V 1 22 E n g l i s h 1 J A N 2010 1 1 in progress is displayed while the load alternator is in motion Press the corresponding number key to activate other functions as desired 1 Ma...

Page 17: ...cycles item from the main menu To do this Press the number key 1 1 1 Mass comparison cycles 2 Move vertically 3 Move horizontally 4 Configure C C 1 0 0 0 0 U L S o f t w a r e V 1 22 E n g l i s h 1 J...

Page 18: ...When the start menu is displayed press the number key that corresponds to the desired number of cycles to start the measurement procedure e g to select 6 A B B A cycles press the number key 2 2 The m...

Page 19: ...cles end the centering procedure Continuous Function With the continuous function option you can repeat complete measurement cycles continuously To stop the continuous function press the reset key Not...

Page 20: ...Move vertically 3 Move horizontally 4 Configure C C 1 0 0 0 0 U L S o f t w a r e V 1 22 E n g l i s h 1 J A N 2010 A new window will displayed select start the mass comparison without start delay ti...

Page 21: ...desired number of cycles to start the mass comparison procedure e g 6 A B B A cycles press the number key 2 2 The message 2 Please confirm by re selecting function is displayed 1 1 A B B A cycle 2 6...

Page 22: ...signation the number of A B B A cycles and the number of weights The load alternator begins by centering the weights The first weight to be tested will centered first as following e g 2 weights on the...

Page 23: ...ent sequence as described above When the 2 2 Please confirm by reselecting function message is displayed press the number key 2 2 and hold it until the message 2 in progress Continuous function is dis...

Page 24: ...e after data is received from the PC so that the next measured values to be received appear in a new line Otherwise the new values will overwrite the former values Data output example CCE1000S L Measu...

Page 25: ...r receiving data on PC from the mass comparator and sorted in an excel file P PC receive P PC Comparator measurement number measurement measurement time A value A time B value B generated 1 Measuremen...

Page 26: ...two question marks will be displayed indicating the system s readiness to receive commands If further data parameters confirmations are required this will be indicated by the appearance of a single q...

Page 27: ...before start A A All 6 comparisons Mass comparison all 6 combinations Only possible when 4 weights are used When required entry before measurement starts R R READ Take weight value from the mass comp...

Page 28: ...he number key 3 3 The load alternator platform is in the centering or lowered position Move the platform to the initial raised position Press the number key 2 2 and then press the key 1 1 while the su...

Page 29: ...eck for and correct unfavorable ambient conditions e g vibrations drafts or air turbulence extreme temperature fluctuations Error message Cause Solution After turning on the mass comparator or during...

Page 30: ...A B B A selected for example the memory can store data from up to 6 complete measurement When the memory is full this is indicated by a crossed out printer symbol in the display The load alternator co...

Page 31: ......

Page 32: ...rius Lab Instruments GmbH Co KG Goettingen Germany All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sa...