14 QuickSeal
®
Cutter
Maintenance
6 Maintenance
6.1 Each Operating Day
Before Use:
t
Inspect blades for wear or damage such as cracks, gouges or chips.
t
Inspect the tool for damage or leaks. If damage is detected, return the tool
to Sartorius Stedim Biotech for inspection.
Skin injection hazard:
Do not use hands to check for oil leaks.
High pressure oil easily punctures skin causing serious injury, gangrene, or death.
If injured, seek medical help immediately to remove oil.
After Use:
t
Wipe all tool surfaces clean with a damp cloth and mild detergent.
Do not use solvents or flammable liquids to clean the cutting tool. Solvents or
flammable liquids could ignite and cause serious injury or property damage.
t
Retract the ram putting the blades into the fully open position.
Place the tool into the carrying case. Store in a cool, dry place.
t
Charge the battery.
6.2 Monthly
Thoroughly clean all surfaces.
6.3 Annually (or after 10,000 cuts)
Return the tool to Sartorius Stedim Biotech for inspection
(see also Chapter “10 Appendix”, page 28).