SARAD Thoron-Scout User Manual Download Page 9

SARAD 

GmbH

 

Thoron-Scout 

 

 
 

Manual_ThoronScout_EN_02-07-15.doc 

9

Statistical Error (for non-mathematicians) 

The radioactive decay is a statistical process. That means, even if the Radon concentration 
is  constant  over  the  time,  the  number  of  decays  N  counted  within  several  intervals  of  the 
same period will be different. N will  vary around the mean value of all considered intervals. 
Considering an infinite number of intervals would lead to an average which one indicates the 
“true” result of N. For a single interval, the value of N will be either below or above the “true” 
value. This observed deviation is covered by the term “Statistical Error”. 
 
Therefore, each serious measurement contains beside the calculated Radon value the error 
band for a stated confidence interval. The commonly used confidence intervals are 1, 2 or 3 
Sigma (

σ

) which refer to a likelihood of 68.3%, 95.45% and 99.73%. 

For example, the correct interpretation of a measured Radon concentration of 780 Bq/m³ with 
a statistical 1

σ

 error of 

±

15% is: 

The real “true” Radon concentration lies with a likelihood of 68.3% within the range from 663 
Bq/m³ (780 Bq/m³ - 15%) to 897 Bq/m³ (780 Bq/m³ + 15%). 

Error Prediction 

The relative statistical error E for a chosen confidence interval of k-Sigma can be predicted 
from the number of detected counts N by the equation: 
 
E[%] = 100% * k * 

(N) / N 

 
The simple consequence is: The higher the number of counts the higher is the accuracy of 
the measurement. From the opposite point of view one could ask: How many counts I have 
to detect to achieve a predefined uncertainty? 
Two items will affect the number of counted decays: The sensitivity of the instrument at the 
one  hand  side  and  the  time  period  used  for  counting  process  (integration  interval)  on  the 
other hand. 
While  the  sensitivity  is  an  instrument  specific  constant,  the  integration  interval  may  be 
expanded  to  the  maximum  acceptable  value  for  the  desired  time  resolution  of  a 
measurement series. 
The relationship between the measured Radon concentration C

Rn

 and the number of counts 

N within an integration interval T is: 
 
C

Rn

 = N / (T * S) 

 
whereby S represents the Sensitivity of the instrument, given in the unit [cts/(min*kBq/m³)]. 
The  sensitivity  using  the  slow  mode  is  double  as  high  as  in  the  fast  mode  (see  chapter 
“Theory  of  Operation”)  and  whenever  the  required  response  time  is  more  than  2  hours  the 
slow mode should be selected. 
For the following examples a fast mode sensitivity of 4 cts/(min*kBq/m³) shall assumed while 
the slow mode sensitivity shall be 8 cts/(min*kBq/m³). 
 
The  first  question  could  be:  What  an  integration  interval  T  have  to  set  to  get  a  statistical 
uncertainty less than 10% at a confidence level of 1

σ

 if the expected Radon concentration is 

200 Bq/m³? 
 
A 1

σ

 error of 10% rquires 100 counts (100%* 1 * 

(100) / 100 = 10%). Using the fast mode, 

the integration interval can calculated by 
 
T(fast) = N / (C

Rn

 * S) = 100 cts / (0.2 kBq/m³ * 4 cts/(min*kBq/m³) = 125 min. 

 
Because  the  required  interval  is  longer  than  2  hours,  the  slow  mode  is  the  better  choice, 
leading to the following result: 

Summary of Contents for Thoron-Scout

Page 1: ...Manual_ThoronScout_EN_02 07 15 doc 1 User Manual Thoron Scout Version 06 2015 SARAD GmbH Tel 49 0 351 6580712 Wiesbadener Stra e 10 FAX 49 0 351 6580718 D 01159 Dresden e mail support sarad de GERMAN...

Page 2: ...5 CARRYING OUT A MEASUREMENT 6 Adjustment of the sampling interval 7 OPERATION MODES 7 Pump 8 Alarm 8 Fast Slow Mode 8 Sniffing 8 DATA HANDLING 8 Data Storage 8 Data transfer RS232 and USB interface 8...

Page 3: ...Minutes This time span defines the minimum achievable response time to a Radon concentration step Now the decay chain is continued by the both beta emitters Pb 214 and Bi 214 followed by another alph...

Page 4: ...tes dislocation during the measurement A display with backlight informs about the actual readings Neither mechanical parts like membrane pumps nor an external power supply are required Therefore use a...

Page 5: ...is placed at the rear panel of the Thoron Scout Changing the batteries will force you to adjust the re set internal real time clock A concerning message appears on the display of the Thoron Scout Sto...

Page 6: ...oron Scout SN 00256 Wait 120 Minutes for first data The actual status and set up information see below may be displayed by pressing the button again If the first interval has been completed five diffe...

Page 7: ...6 19 12 34 ALM 10000Bq m INTVL BUZZ OFF TSC 00256 To finish a measurement series keep the Toggle button down and wait for at least four beeps from the buzzer If the button is released the sample will...

Page 8: ...should be cleared after successful data transmission and storage on hard disk This will save time during the next transmission and avoids redundant data storage Each data record is stored after compl...

Page 9: ...e accuracy of the measurement From the opposite point of view one could ask How many counts I have to detect to achieve a predefined uncertainty Two items will affect the number of counted decays The...

Page 10: ...her than the lower limit of the error band of reading 1 Because the true value could be placed within 1350 Bq m and 1469 Bq m the variation of both readings is not statistical significant Example 2 Re...

Page 11: ...han 16 and therefore the statistical fluctuations have to be derived by the Poisson distribution The stated confidence interval gives the probability that the detected number of decays within the inte...

Page 12: ...erval 1 255 Minutes adjustable Memory 2047 data records Internal volume chamber approx 60 ml Power supply Battery operation 30 days AC DC adapter 4 5V 0 5A PC Interface Serial USB or RS232 9600 Baud 8...

Reviews: