4. After Green LED glows, turn ’CA’ pot slowly anti-
clockwise until Red LED just goes ON.
5. Empty the tank up-to lower level point.
6. Throw the calibration switch ’CAL’ to ’
∆
S’ position.
Red LED glows.
7. Turn ’
∆
S’ pot slowly clockwise until green LED just
goes ON.
8. Throw calibration switch ’CAl’ to ’
∆
S’ position.
CAUTION - Do not move ’CA’ or ’
∆
S’ pot here after.
14
Calibration Procedure For
Fail-Safe Low
•
After following similar instructions 4 (a) to (g) throw
the fail safe select switch ’FSS’ to fail safe ’LO’ po-
sition.
The instrument is now calibrated.Ensure the calibration
is correct by observing the filling and emptying cycle.