- 83 -
37
SPECIFICATIONS
Name
Ultra-Low Temperature Freezer
Model
MDF-U74V
MDF-U74VC (For U.S.A. only)
External dimensions
W1010 x D870 x H2010 (mm)
Internal dimensions
W870 x D600 x H1400 (mm)
Effective capacity
728 L
Exterior
Painted steel
Interior
Painted steel
Outer door
Painted steel
Inner door
ABS resin panel with stainless frame, 2 doors
Shelf
Stainless steel, 3 shelves (adjustable)
Inner dimension; W848 x D533 (mm), Load; 50 kg/shelf
Access port
17 mm diameter, 3 locations (back, bottom x 2)
Insulation
Vacuum insulation panel + Rigid polyurethane foamed-in place
Compressor
High stage side; Hermetic type, Output; 1100 W
Low stage side; Hermetic type, Output;1100 W
Evaporator
Tube on sheet type
Condenser
High stage side; Fin and tube type, Low stage side; Shell and tube type
Refrigerant
High stage side; R-407D, Low stage side; R-508
Temperature controller
Microcomputer control system
Temperature display
Digital display
Thermal sensor
Platinum resistance (Pt 1000
ǡ
)
Alarm
High temp. alarm, Low temp. alarm, Power failure alarm
Door alarm, Filter alarm
Remote alarm contact
Allowable contact capacity: 30 VDC, 2 A
Battery
Nickel-metal-hydride battery, 6 VDC, 1100 mAh, Auto-recharge (5HR-AAC)
Accessories
1 set of key, 1 scraper
Weight 346
kg 351
kg
Voltage booster
None
Built-in
Optional component
Inventory rack (IR-220U, IR-224U), Independent inner door (MDF-7ID),
Automatic temperature recorder (MTR-G85)*,
Interface board (MTR-480), LAN interface board (MTR-L03),
Data acquisition system (MTR-5000),
Back-up system (CVK-UB2, CVK-UB2(I)): LCO
2,
(CVK-UBN2): LN
2
Note
:
• Design or specifications will be subject to change without notice.
• The battery for power failure alarm is an article for consumption. It is recommended that the
battery will be replaced about every 3 years. Contact Sanyo sales representative or agent at the
time of replacement of the battery for recycling.
• Fan motors are expendable supplies. Replace them for about every 6 years. Contact Sanyo
sales representative or agent at the time of replacement of the fan motor.
*: Power source of the automatic temperature recorder shall be 220 V.