TROUBLE SHOOTING
If the unit malfunctions, check out the following before calling for service.
Malfunction Check/Remedy
The chamber is not cooled
at all
㩵
The circuit breaker of power source is active.
㩵
The voltage is too low (In this case, call an electrician).
㩵
The power is not supplied.
㩵
The breaker is free.
㩵
The large amount of articles (load) is stored in the chamber at one
time.
The cooling is poor
㩵
The ambient temperature is too high.
㩵
The latch of inner door is not closed completely. The outer door
is not closed firmly. (The frost or ice between the cabinet and door
gasket possibly prevents door seal.)
㩵
The air intake vent is blocked.
㩵
The condenser filter is clogged. Always clean the filter when the
filter check lamp is lit.
㩵
The door is not shut tightly.
㩵
The inner door is not installed correctly.
㩵
The set temperature in the controller is not set properly.
㩵
The grille is blocked out.
㩵
The filter is clogged.
㩵
The freezer is in the direct sunlight.
㩵
There is any heating source near the freezer.
㩵
A rubber cap and insulation are not set correctly.
㩵
You put too many unfrozen articles into the chamber.
Alarm test key cannot
actuate the alarm
㩵
The alarm is activated only when the power switch is ON.
㩵
When only the buzzer or only the alarm is actuated by the alarm test
key, the unactuated part is out of order, and must be replaced.
Note:
If the malfunction is not eliminated after checking the above items, or the malfunction is not shown in the
above table, contact Sanyo sales representative or agent.