29
−
10.
10.
10.
10.
Function mode
Functions are as follow.
F00
:
To return to PV display
F01
:
To change temperature of high temperature alarm
F02
:
To change temperature of low temperature alarm
F03
:
To change temperature of high temperature alarm
(Applied for lower compartment of MDF-U537D)
F04
:
To change temperature of low temperature alarm
(Applied for lower compartment of MDF-U537D)
F05
:
To set compressor delay start time
F06
:
To input service code (384)
F07
:
To adjust zero for temperature sensor
(Upper compartment sensor for MDF-U537D)
F08
:
To adjust zero for temperature sensor
(Not available for MDF-U333)
(Solenoid valve control sensor for MDF-U537)
(Lower compartment sensor for MDF-U537D)
F09
:
Not available
F10
:
Not available
F11
:
Not functioned. Return to PV display.
F12
:
To display temperature of temperature sensor (per 0.1
℃
)
F13
:
To display temperature of MDF-U537/U537D compressor protect sensor (per 1
℃
)
F14
:
To display temperature of temperature sensor (per 0.1
℃
)
(Solenoid valve control sensor for MDF-U537)
(Lower compartment sensor for MDF-U537D)
F15
:
Not available
F16
:
Not functioned. Return to PV display.
F17
:
To set or check model code
model code
001
:
MDF-U333
model code 002
:
MDF-U537
model code 003
:
MDF-U537D
F18
:
Preset value for second offset performance: -26
℃
(for MDF-U537/U537D)
F19
:
Not functioned. Return to PV display.
F20
:
Not functioned. Return to PV display.
F21
:
To set communication ID
(
000
~
255
)
F22
:
To set communication mode
F23,24,26
~
29
:
With input each function code, press SET key to sound buzzer 1 second
continuously to maintain input status.
F25
:
To set alarm ring back time
Setting value: 000,010,020,030,040,050,060
Setting method
:
:
:
:
In PV display, keep key pressing over 5 seconds to enter function mode and will be
displayed “F00”. Press key to input required function code then press
SET key again.
F00
:
Not functioned. In F00 displayed, press SET key to return PV display.
Summary of Contents for MDF-U333
Page 12: ... 9 Dimensions MDF U333 ...
Page 13: ... 10 Dimensions MDF U537 ...
Page 14: ... 11 Dimensions MDF U537D ...
Page 18: ... 15 Refrigeration circuit principle diagram MDF U333 ...
Page 19: ... 16 Refrigeration circuit welding points MDF U333 ...
Page 20: ... 17 Refrigeration circuit principle diagram MDF U537 ...
Page 21: ... 18 Refrigeration circuit welding points MDF U537 ...
Page 22: ... 19 Refrigeration circuit principle diagram MDF U537D ...
Page 23: ... 20 Refrigeration circuit welding points MDF U537D ...
Page 37: ... 34 Wiring diagram ...
Page 38: ... 35 ...
Page 39: ... 36 ...
Page 40: ... 37 Circuit diagram Main PCB ...
Page 41: ... 38 Components on PCB Main PCB ...
Page 52: ... Memorandum ...