REPLACEMENT OF BATTERY
The battery for power failure alarm is an article for consumption. The battery life is approximately 3
years. The buzzer can not be activated at the power failure and the stored items may be influenced if
the battery is left as it is for more than 3 years. It is recommended that the battery is replaced ahead of
time.
For the replacement of the battery, contact Sanyo sales representative or agent.
Location of a lead storage battery
This unit is provided a lead storage battery for the power failure warning device. The battery is located
in the battery box inside the side table. (Fig. 1)
The high voltage components are enclosed in the battery box. The cover should be removed by
a qualified engineer or a service personnel only to prevent the electric shock.
Removal of lead storage battery
1.
Turn off the power switch and disconnect the power supply plug.
2.
Lift the handle of the side table to the top as shown in the fig. 2, and open the door of the side table.
Remove 6 screws of the battery box cover, and remove a battery box cover.
3.
Remove 4 screws which fix a battery fitting on the side table. (Fig. 3)
4.
Remove the connector of the battery, and take out a battery from the side table, and remove 4 screws
of the fitting which fixes a battery (4 pcs). (Fig. 4)
Handling of the lead storage battery:
Cover the battery terminal with an insulating tape to avoid the short circuit. Then follow the procedure
for recycling or proper disposal.
Fig.
3
Battery bracket
Screw
Side table
Fig.
2
Battery box
cover
Screw
Screw
Battery bracket
Fig.
1
Battery
Side table
Connector
Battery
Fig.
4
29