15. Lamp, display and buzzer specifications
Lamp operation:
<Control
PCB>
DP1:
DP2:
DP3:
DP6:
<Display PCB>
DP51:
DP53:
DP54:
DP55:
Examples of display:
Orange
lamp
Goes off: High/low temp. alarm (15min. delay), sensor error,
power failure
Lit
: Not in alarm condition
Green
lamp
Goes off: L side compressor turns off. (normal condition)
Lit
: L side compressor turns on.
Red
lamp
Goes off: Capillary heater turns off. (normal condition)
Lit
: Capillary heater turns on.
Green
lamp
Goes off: H side compressor turns off. (normal condition)
Lit
: H side compressor turns on.
Red
lamp
Blinks : High/low temp. alarm (without delay), or sensor failure,
or power failure
Orange
lamp
Lit
: In filter alarm
Red
lamp
Lit
: Door ajar
Orange
lamp
Lit
: Battery accumulation time is reached to about 3 years
Blinks : Fan motor accumulation time is reached to about 6 years
Decimal point of chamber temp. -80.0
Chamber temp. -79.5
㷄
Set temp.
-80.0
㷄
Sensor
offset
㩷 㩷 㩷
-5.0
Operation monitoring
㩷 㩷
LCP
Function
㩷 㩷 㩷
F03
Error
㩷 㩷 㩷 㩷
E01
Service code
㩷 㩷 㩷
384
Set value
㩷 㩷 㩷
004
Accumulation time
㩷 㩷
8H timer 135
Key Lock
㩷 㩷 㩷
L_0
Buzzer
tone:
Alarms
Key operation
Set value memory
Out of settable range
Filter blockage
Intermittent
tone
Click tone
Click tone
1 second continuous tone
Intermittent tone