3. pH Calibration 

3.1. Before first use or unused for a long time, fill certain 3M KCl solution in the calibration bottle, 

soak probe for at least 15 minutes. Pay attention that, soaking solution little higher that pH bulb 

will be enough. Users can make this solution by themselves. Preparation: dissolve 25g KCl in 100 

ml of pure water. 

 
3.2. Rinse the probe in pure water and dry it.   

                                                                                 
3.3. Fill certain amount (about half volume of calibration bottle) of pH7.00 and 

pH4.00 buffer solution in separate pH calibration bottles;   

 
3.4. Long press          to enter calibration mode; Short press          to exit. 

 
3.5. Dip the probe in pH7.00 buffer solution and stir gently, leave to Stabilize 

  (LCD displays        as shown in the picture on the right). Short press          to 

  complete 1-point calibration. The tester then returns to measurement mode and calibration icon           

          appears on the LCD.   

 
3.6. For multi-point calibration (either 2 or 3 point), rinse the probe in clean water and repeat   

step 3.5 using 4.01pH and/or 10.01pH calibration solutions. After completion of each point, the 

calibration icons          (for 4.01pH) and          (for 10.01pH) will appear on the LCD near the 

     

.

 

 

4. pH Measurement 

Short press          to turn on the tester. Rinse the probe in pure water and dry it. Dip the probe in 

sample solution, stir gently, leave it to stabilize.

 

Get readings after LCD displays

       

 

5. Special Notes 

5.1. Tester can perform 1~3points automatic calibration. But pH7.00 or pH6.86 buffer solution 

must be used to make 1st point calibration. Then use other buffer solution to conduct 2nd or 3rd 

point calibration. Tester will recognize 5 kinds of pH buffer solutions. Refer to the following chart:   

 

Calibration 

USA Series 

NIST Series 

Indication icon 

Recommended     

1-point     

7.00 pH 

6.86 pH 

 

Accuracy ≥ 0.1pH 

2-point     

7.00 pH, 

4.01 pH or 1.68pH 

6.86 pH, 

  4.01 pH or 1.68pH 

 

Range: 

7.00pH 

7.00 pH, 

10.01pH or 12.45pH 

6.86 pH, 

9.18 pH or 12.45pH 

 

Range: 

7.00pH 

3-point     

7.00pH, 4.01 or 1.68pH, 

10.01 or 12.45pH 

6.86pH, 4.01 or 1.68pH, 

9.18 or 12.45pH 

 

Range: 0~14.00pH 

Diagram - 2 

 

Summary of Contents for PC5

Page 1: ...PC5 pH Conductivity Tester pH Conductivity TDS Salinity Temp Operation Manual Shanghai San Xin Instrumentation Inc ...

Page 2: ...etting short press to change parameter Unidirectional 1 Long press to enter calibration mode 2 In calibration mode short press to confirm calibration 3 When measured value is locked short press to unlock Pull the battery cap up Push the battery cover along to the direction of arrow Open the battery cover Insert the batteries 4 AAA and ensure the batteries polarity is as per drawing Close the batte...

Page 3: ...r 3 point rinse the probe in clean water and repeat step 3 5 using 4 01pH and or 10 01pH calibration solutions After completion of each point the calibration icons for 4 01pH and for 10 01pH will appear on the LCD near the 4 pH Measurement Short press to turn on the tester Rinse the probe in pure water and dry it Dip the probe in sample solution stir gently leave it to stabilize Get readings after...

Page 4: ...ently leave to Stabilize LCD display as shown in the picture on the right Short press key to complete 1 point calibration The tester then returns to measurement mode and indication icon will appears on the LCD 6 5 For multi point calibration either 2 or 3 point rinse the probe in clean water and repeat step 6 4 using 84uS standard solution and or 12 88mS standard solution After completion of each ...

Page 5: ...n 1 72 C 5 sulfuric acid solution 0 96 C Dilute ammonia solution 1 88 C 8 4 TDS and conductivity is linear related and its conversion factor is 0 40 1 00 Adjust the factor in parameter setting P5 The factory default setting is 0 71 Salinity and conductivity is linear related and its conversion factor is 0 5 The tester only needs to be calibrated in conductivity mode then after calibration the mete...

Page 6: ...n reading is stable for more than 10 seconds the tester will lock the value automatically and HOLD icon will display on LCD Press key to cancel automatic hold c Backlight P3 Off turn off backlight On turn on backlight 1 backlight will last for 1 minute d Temperature compensation factor P4 Adjust temperature compensation factor between 0 00 4 00 Factory default setting is 2 00 Details refer to clau...

Page 7: ...strument does not work well in calibration or measurement Calibrate and measure again after recover the instrument to factory default status 10 Technical Specifications pH Measuring range 2 00 16 00pH Resolution 0 01pH Accuracy 0 01pH Calibration points 1 3points automatic calibration Automatic temperature compensation 0 50 Cond Measuring range 0 200 0μS 0 2000μS 0 20 00mS Resolution 0 1 1μS 0 01m...

Page 8: ... screw on the screw ring 13 Warranty We warrant this instrument to be free of defects in parts and workmanship for two years from date of shipment a six month limited warranty applies to probes This warranty does not apply to defects resulting from actions of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Warra...

Reviews: