![Sanwa PC710 Instruction Manual Download Page 25](http://html.mh-extra.com/html/sanwa/pc710/pc710_instruction-manual_1198168025.webp)
― 46 ―
― 45 ―
To make AC current measurement (ACA), set the function
selector to
and press the SELECT button to select m /Hz,
then press the RANGE button to set 600.0mV range.
③
Set the range selector knob on the current probe to the 20A
range or 200A range.
*Before making DC current measurement, turn the Center
Adjuster knob to make the reading zero.
④
Open the clamp jaws of the clamp probe and clamp the wire
to measure.
⑤
Multiply the reading by 0.1 for 20A range, and read the
display directly for 200A range.
Note:
・
Current exceeding 20A or 200A cannot be measured.
(Do not measure such high current even though the display
works.)
・
Try to put the wire to measure in the center of the clamp jaws
as possible.
5-10-3 DC Clamp probe: CL-33D
(Max. measurable current 300A dc)
1) What to measure
Current of automotive electric circuits, consumption current of
DC equipments, etc.
2) Measuring ranges
30A and 300A
3) Measuring procedure
①
Connect the red plug of the current probe to the V measuring
terminal and the black one to the COM terminal.
②
Set the function selector to
and press the SELECT
button to select m , then press the RANGE button to set
600.0mV range.
③
Set the range selector knob on the current probe to the 30A
range or 300A range.
*Before making DC current measurement, turn the Center
Adjuster knob to make the reading zero.
④
Open the clamp jaws of the clamp probe and clamp the wire
to measure.
⑤
Multiply the reading by 0.1 for 30A range, and read the
display directly for 300A range.
Note:
・
Current exceeding 30A or 300A cannot be measured.
(Do not measure such high current even though the display works.)
・
Try to put the wire to measure in the center of the clamp jaws as
possible.
5-10-4 Temperature probe: T-300PC
1) What to measure
Temperature of liquid, solids, gas, and etc.
Note:
To make temperature measurement, connect the temperature
probe to the PC710 connected to the PC on which sanwa's
software PC Link7 is installed and running.
2) Measuring range
-50
〜
300
℃
'00UDQJHN
3) Measuring procedure
①
Connect the red plug of the temperature probe to
measuring terminal and the black one to the COM terminal.
②
Set the function selector to
and press the SELECT button
WRVHOHFW>@
③
3UHVVWKH5$1*(+2/'EXWWRQWRVHWN
④
Apply the thermocouple to the object to measure.
⑤
Read the measurements on the information window of the
PC Link7.
⑥
Remove the thermocouple from the object.