OVERVOLTAGE CATEGORY
• Equipment of CAT
is equipment for connection to circuits
in which measures are taken to limit the transient
overvoltages to an appropriate low level.
Note:
Examples include protected electronic circuits.
• Equipment of CAT
is energy-consuming equipment to be
supplied from the fixed installation.
Note:
Examples include household, office, and laboratory
appliances.
• Equipment of CAT
is equipment in fixed installations.
Note:
Examples include switches in the fixed installation and
some equipment for industrial use with permanent
connection to the fixed installation.
• Equipment of CAT
is for use at the origin of the installation.
Note:
Examples include electricity meters and primary over-
current protection equipment.
8-2 Measurement Range and Accuracy
Accuracy assurance range: 23
5 , 75%RH MAX.
No condensation.
TRUE RMS ACV & ACA accuracies are specified from 5 % to 100
% of range or otherwise specified, crest factor <3:1 at full scale &
<6:1 at half scale, and with frequency component within the
specified frequency bandwidth
DC Voltage
— 34 —
RANGE
Accuracy
50.00 mV
0.12% rdg + 2dgt
500.0 mV
0.06% rdg + 2dgt
5.000V, 50.00V, 500.0V, 1000V
0.08% rdg + 2dgt
NMRR:>60dB @ 50/60Hz
CMRR:>120dB @ DC, 50/60Hz, Rs=1k
Ω
Input Impedance: 10M
Ω
, 16pF nominal (44pF nominal for 50mV
& 500mV ranges)