34
5-2 Error rate
Error rate is represented as the percentage ratio of reading M on the unit under measurement (the
PC20TK) minus true value T (reading on the standard) to true value T.
M−T
= - × 100 (%)
T
5-3 Example of completing the table for test results
Table for Test Results
Model :
PC20TK
Tested on :
March 10, 2008
Serial No. :
2031234
Test conditions : Temperature
23
°C ; relative humidity
60
%
: Frequency of AC voltage and current
50
Hz
Tested by:
Mr. Saburo Sanwa
Functional mode
/range
Reading M on
PC20TK
(displayed value)
Reading T on
standard
(true value)
Error rate
Accuracy range
DC 400 mV
380.0 mV
378.01 mV
+0.53 %
376.0~384.0 mV
DC 4 V
3.800 V
3.7752 V
+0.66 %
3.760~3.840 V
DC 40 V
38.00 V
37.732 V
+0.71 %
37.60~38.40 V
DC 400 V
380.0 V
378.21 V
+0.47 %
376.0~384.0 V
DC 750 V
600 V
594.1 V
+1.00 %
592~608 V
AC 4 V
3.800 V
3.8338 V
−0.88 %
3.734~3.866 V
AC 40 V
—
—
—
—
AC 400 V
380.0 V
384.04 V
−1.05 %
373.8~386.2 V
AC 750 V
—
—
—
—
DC 400 µA
380.0 µA
377.52 µA
+0.66 %
374.1~385.9 µA
DC 4000 µA
—
—
—
—
DC 40 mA
38.00 mA
37.631 mA
+0.98 %
37.41~38.59 mA
DC 400 mA
—
—
—
—
AC 400 µA
—
—
—
—
AC 4000 µA
3800 µA
3815.1 µA
−0.39 %
3719~3881 µA
AC 40 mA
—
—
—
—
AC 400 mA
380.0 mA
383.73 mA
−0.97 %
371.9~388.1 mA
40 nF
—
—
—
—
400 nF
—
—
—
—
4 µF
2.200 µF
2.1049 µF
+4.52 %
2.040~2.360 µF
40 µF
10.00 µF
9.761 µF
+2.45 %
9.24~10.76 µF
100 µF
—
—
—
—
Ω
400
Ω
200.0
Ω
198.56
Ω
+0.72 %
196.5~203.5
Ω
Ω
4 k
Ω
—
—
—
—
Ω
40 k
Ω
20.00 k
Ω
19.901 k
Ω
+0.50 %
19.65~20.35 k
Ω
Ω
400 k
Ω
—
—
—
—
Ω
4 M
Ω
1.000 M
Ω
0.9906 M
Ω
+0.95 %
0.980~1.020 M
Ω
Ω
40 M
Ω
—
—
—
—