8-1 General Specifications
Operation method
Delta-sigma modulation
AC sensing method
Average detection
LCD
9.99/99.9/110 (MΩ), 610.0 (Voltage),
9999 (Resistance)
Sampling rate
Approx. 2 times / sec.
Over-range
indication
"OL" on the numerical part and beep
The LED level meter blinks.
Range selection
Automatic selection
Polarity switching
Automatic switching ("-" will be indicated
as necessary.)
Low battery
indication
Lower than approx. 4.0 V
~
4.3 V (depending on
the function):
indicated
Operating conditions Altitude:
≦
2,000 m, Pollution degree: II,
Indoor use only
Operating temperature/
humidity
0
℃
to 40
℃
: 90 %RH (Max.) non-
condensing
Storage temperature/
humidity
-10
℃
to 50
℃
: 90 %RH (Max.) non-
condensing (with battery removed)
Power source
Alkaline battery 1.5 V (LR03) 4 pieces
Auto Power Off
Power-off approx. 10 min. after the last operation.
Max. Power
consumption
Approx. 2.5 VA (when 500 V, 0.5 MΩ
measurement)
Possible
measurement times
1,000 times (when 500 V, 0.5 MΩ, ON: 5
sec./ OFF: 25 sec.)
Dimension, Mass
H
139 x
W
91 x
D
29 Approx. 230 g (with batteries)
Manufacturing year
Most significant 2 digits of the serial
number on the back of the instrument
show the lower 2 digits of the
manufacturing year (20xx).
IP Protection rating
IP30 (with measurement probe)
Safety Compliances
IEC61010-1
CAT.III 300 V, II 600 V
IEC61010-2-030, IEC61010-2-033
IEC61557-1 and 2, IEC61010-031
EMC, RoHS
IEC61326-1(EMC), EN50581(RoHS)
Accessories
Instruction manual, Measurement probe
(TL-561), Alligator clip (CL-561), Carrying
case (C-DG3a)
(The instrument usually accepts Ni-MH rechargeable batteries,
and may not work without the low battery indication caused by
the discharge characteristic of the batteries. We recommend you
have spare batteries with you just in case.)
/ distributor without above information will be returned to
the customer.
Note:
1) Prior to requesting repair, please check the
following:
Capacity of the built-in battery, polarity of installation
and discontinuity of the test leads.
2) Repair during the warranty period:
The failed meter will be repaired in accordance with the
conditions stipulated in 7-1 Warranty and Provision.
3) Repair after the warranty period has expired:
In some cases, repair and transportation cost may
become higher than the price of the product. Please
contact Sanwa authorized agent / service provider in
advance.
The minimum retention period of service functional
parts is 6 years after the discontinuation of
manufacture. This retention period is the repair warranty
period. Please note, however, if such functional parts
become unavailable for reasons of discontinuation of
manufacture, etc., the retention period may become
shorter accordingly.
4) Precautions when sending the product to be
repaired:
To e n s u r e t h e s a f e t y o f t h e p r o d u c t d u r i n g
transportation, place the product in a box that is larger
than the product 5 times or more in volume and fill
cushion materials fully and then clearly mark “Repair
Product Enclosed” on the box surface. The cost of
sending and returning the product shall be borne by
the customer.
7-3 SANWA web site
http://www.sanwa-meter.co.jp
E-mail: [email protected]
[8] SPECIFICATIONS