Sanosil Q-Jet Compact 

 

 

 

 

 

2.2

 

Intended Usage

 

The device is a nebulizer to be used for indoor applications only. 

The device is a nebulizer to be used for indoor 

applications with Sanosil disinfectants only maximal hydrogen Peroxide content of 7.5%.

 

Its purpose is to transform the used disinfection liquid into smallest particles and to nebulize them into a room, thus 
leading to the consistent disinfection of all uncovered surfaces (3D-disinfection) inside the room.

 

This device is intended for usage in hospitals, clinics, public transports, etc. as well as in residential-, commercial- and in- 
dustrial areas. It is an additional measure following regular cleaning and disinfecting room surfaces. Pre-cleaning protocol 
is defined by the respective healthcare facility. Ensure surfaces are dry before nebulizing in order to maximize the impact.

  

 

 

 

 
 

WARNING

 

Risk of Damage!

 

This device nebulizes chemical substances!

 

The area to be disinfected may only contain items necessary for its use. 
Objects made of delicate organic materials, in particular antiques, paintings and precious artifacts 
are to be removed!

 

 

2.3

 

Inappropriate Usage

 

Do always and only use suitable Sanosil disinfectants with a maximal hydrogen Peroxide content of 7.5%. The usage 
of any other disinfectant or accessories which are not specifically approved by Sanosil Ltd. is against the rules herein 
and can lead to heavy injuries as well as to damaging the device. The device is not meant for use as a stand-alone 
disinfection process but rather as an addition to regular cleaning and disinfection. 

 

 

 

DANGER

 

Risk of Injuries!

 

Danger of explosion

!

 

Do not use any flammable liquids

Do always and only use Sanosil disinfection liquids.

 

 

2.4

 

Danger and Caution Warnings

 

 

The device must be used in full compliance with this Operation and Maintenance Manual.

 

 

It is strictly forbidden to use the device in an atmosphere with danger of explosion or flammability, in dusty areas 

and in areas with the presence of people or animals. It is an imperative to prevent people from entering 
intentionally or inadvertently the room during the nebulization process and during the waiting time thereafter.

 

 

The device can only be operated using the original power cord or a country-specific power cord with earthing wire 

and plug. Before making electrical connections carefully check that the voltage supplied is compatible with the 
voltage marked on the 

device’s

 data plate. Any electrical installation work must be carried out by qualified 

personnel.

 

 

The power socket must be easily accessible.

 

 

Risk of injuries when using not authorized accessories!

 

 

Therefore, do only use original accessories of Sanosil Ltd.

 

 

Place device on a stable, flat, dust-free, not slippery, horizontal surface which can appropriately support its weight.

 

 

Before removing the rear panel, always disconnect the device from the power supply.

 

 

When there is liquid inside the tank, do not tilt the device; it could leak from the hydraulic circuit.

 

 

Read carefully all indications stated on the label of the Sanosil disinfectant.

 

2.5

 

Authorized Persons

 

The device shall only be used by adults who have been carefully instructed on how to operate it and are well informed on 
possible dangers and risks incurred when using the device.

 

The users must always have a copy of this Operation Manual at their disposal when operating the device and have 
read and understood all specific indications stated on the label of the Sanosil disinfectant used. 

 

 

Summary of Contents for Q-Jet Compact

Page 1: ...Sanosil Q Jet Compact 1 Operation and Maintenance Manual Sanosil Q Jet Compact Des M 56en Compact V2 English...

Page 2: ...8 3 1 Technical Data 8 3 2 Parts and Functions 8 3 3 Package Contents 8 3 4 Q Jet Compact Accessories 8 3 5 Spare Parts 8 3 6 Recommended Equipment 8 3 7 Trolley optional 9 4 Getting Started and Oper...

Page 3: ...ries Low risk Disregard can lead to medium or light injuries Disregard can damage thedevice Gives good advice for operating thedevice Further symbols will be explained thereafter 2 Safety DANGER Instr...

Page 4: ...Sanosil Q Jet Compact 4 2 1 CE Declaration of Conformity manufacturer...

Page 5: ...le liquids Do always and only use Sanosil disinfection liquids 2 4 Danger and Caution Warnings The device must be used in full compliance with this Operation and Maintenance Manual It is strictly forb...

Page 6: ...an as needed basis depending on the volume of traffic and use Users should familiarize themselves with Q Jet operation by reading the Q Jet User Manual prior to first use It is essential that Q Jet u...

Page 7: ...rical plug or parts are wet 8 Use only a Sanosil supplied electrical power cord 9 Take care to prevent water or other fluid from entering the device Should this occur allow to completely dry before us...

Page 8: ...ent 0 20ppm for example Dr ger X am 5100 Gloves 3 Product Description 3 1 Technical Data Powerrequirements 220 240 V 50Hz Amperage max 5 A Operatingtemperature 5 C to 40 C Engine Power 800 W Engine ro...

Page 9: ...place the device on top of trolley secure the device with the screw on the trolley WARNING Risk of crushing Hands and fingers can be crushed between device and trolley Do always wear device on handle...

Page 10: ...ctrical connections carefully check that the voltage supplied is compatible with the voltage marked on the dataplate Insert power cord 2 in IEC power socket 4 3 Filling Remove tank cap 6 Fill disinfec...

Page 11: ...x Width x Height of the room in meters and round it up to the next cubic meter Check relative humidity should not be lower than 50 Establish nebulization time room s volume x desired concentration 30...

Page 12: ...the waiting time the room can be prepared for re entry please see 4 5 3 Shortening of waiting time after a minimal waiting time of 120 minutes place a strong dehumidifier in the middle of the room and...

Page 13: ...n equipment if the H2 O2 Value isbelow the local maximum permissible value specified by the authorities Tip ON OFF 3 to switch off device INDICATION For transportation as well as for longer storage th...

Page 14: ...eneral Visual control Checking operationality daily Components status of device s components integrity of components fluid loss sealings caps monthly Documentation Verification that all documentation...

Page 15: ...gine temperature Air intake filter toodirty Check engine switch off and on Set up device once more Clean or change filter Liquid is not sprayed Tank is empty Nozzle is clogged Fill liquid Remove obstr...

Page 16: ...ice should be noted please proceed as follows Download Customer Complaint Form from https www sanosil com de q jet and fill it carefully Call CleanService Group Ltd to obtain a Return Merchandise Auth...

Reviews: