11. Technical data:
Type:
SBM 07
Measuring method:
non-invasive, oscillometric blood pressure measurement at the wrist,
for wrist circumferences of 13.5 – 22 cm
Display range:
0 – 299 mmHg
Measuring range:
systolic: 40 – 250 mmHg / diastolic: 40 –180 mmHg /
pulse: 40 –199/min
Accuracy of the pressure reading:
systolic ± 3 mmHg / diastolic: ± 3 mmHg /
pulse: ± 5% of the displayed value
Measuring uncertainty:
(max. permissible standard deviation according to clinical testing in
accordance): systolic: 8 mmHg / diastolic: 8 mmHg
Power supply:
2 x 1.5 V Micro (Alkaline Type LR 03)
Classification
Application part Type BF
Ambient conditions:
+10 °C – +40 °C max. 85% atmospheric humidity
Storage/transport:
-10 °C – +60 °C max. 95% atmospheric humidity
Key to symbols:
Application part Type BF
Important! Read the operating instructions