OPTIONAL ACCESSORIES 

1. 

ArmorLite

TM

 Lamp Safety Shield:

 A protective coating applied to 

STER-L-RAY

®

 Germicidal Ultraviolet Lamps, ensuring 

protection by eliminating the dangers associated with fragments of broken glass and mercury contamination.

2. 

Zenith

 

UV-C Detector:

 A sensitive, hand-held, self-contained battery operated ultraviolet meter that can be used for:

•  Monitoring UV-C Germicidal Lamp Intensity & Aging
•  Measuring UV-C Germicidal Fixture Leakage.

3. 

Promate

TM

 Safety Glasses: 

Safety eyewear 

MUST

 be used as general-purpose safety protection and for additional shielding 

from ultraviolet rays.

4. 

Promate

TM

 Face Shield: 

Lightweight visor with adjustable headgear provides eye and face protection from germicidal 

ultraviolet rays. 

5. 

SENTINEL

®

 

Lamp Operation Indicator:

 Monitors individual lamp operation, in any location up to 100 feet from the 

SANITAIRE

®

 

Room Air Sanitizer. Cables are also available in 25, 50, and 100 foot lengths.

For more information and purchasing contact our offices, weekdays between 8:30 am and 5:00 pm Eastern time, at:
Atlantic Ultraviolet Corporation

®

, 375 Marcus Boulevard, Hauppauge, New York, 11788  

Tel: 631.273.0500 • Fax: 631.273.0771 • Email: [email protected] • Website: Ultraviolet.com – BuyUltraviolet.com

TROUBLESHOOTING 

 

  

Always disconnect power to the unit before performing any service or maintenance.

 Use personal protection equipment, such long sleeves with no gaps between cuffs and gloves, and 

ultraviolet resistant face shield, if troubleshooting 

SANITAIRE

®

.

I

MPORTANT:

 This unit is to be serviced 

ONLY

 by qualified, and appropriately trained and/or licensed personnel.

Table 1 – Troubleshooting

Problem

Possible Cause

Corrective Action

Unit not operating.

No electrical power.

Verify that the unit is connected to a live power source.

Power connection to fixture is loose 
or disconnected.

Verify power connection to fixture is fully engaged.

Internal blower(s) not 
functioning.

Loose power connection.

Confirm connection to blower; connections are tight 
and make full contact.

Faulty blower motor.

Replace blower motor.

Two 

STER-L-RAY

®

 

germicidal lamps 
not operating

Surelite

TM 

 

Ballast not functioning.

Swap both lamps, with known good lamps. 
If good lamps still do not operate, locate 
corresponding ballast and replace.

STER-L-RAY

®

 

Germicidal lamp 

not operating.

Swap both lamps, with known good lamps.

STER-L-RAY

®

 

Germicidal lamp not 

properly seated.

Confirm lamps are installed correctly and are seated 
making full contact with lamp socket.

All 

STER-L-RAY

®

  

germicidal lamps 
not operating.

STER-L-RAY

®

 

Lamps not installed.

Install lamps.

Interlock switch open.

Confirm access door is closed and latched.

Interlock switch faulty.

Replace interlock switch.

If the above measures fail to correct the problem(s), contact our Application Specialists. Contact options can be found in the 

“User Assistance”

 section of this manual, located on the back page.

Summary of Contents for RSCS280A

Page 1: ...ntic Ultraviolet Corporation Read and Follow All Safety Instructions Save These Instructions Installation Operation Maintenance Model RSCS280A Extensive Product Information Available at Ultraviolet com BuyUltraviolet com Owner s Manual Owner s Manual ...

Page 2: ...oom Air Sanitizer Model RSCS280A Questions that are not specifically answered by these instructions should be directed to the Factory Atlantic Ultraviolet Corporation takes all possible precautions when packaging equipment to prevent damage Carefully inspect and report all damages immediately Do not install damaged equipment Follow all instructions on any labels or tags Carefully inspect all packi...

Page 3: ...DANGER WARNING No 00 0195A2 UV ULTRAVIOLET RADIATION Protect eyes and skin from direct and reflected radiation DANGER Turn off and disconnect power before servicing HAZARDOUS VOLTAGE No 00 0197A2 AIR WATER Part No 00 0195A2 2 75 x 1 5 Part No 00 0197A2 2 75 x 1 5 AIR Part No 00 0198A1 1 25 x 3 This fixture is designed for use with germicidal ultraviolet lamps and must be installed in compliance wi...

Page 4: ...sure constructed of stainless steel Stainless steel construction provides strength durability and will withstand most harsh environments An interlock safety switch The interlock switch interrupts power to the sanitizer when the access door is opened from the lamp exposure chamber Principle of Operation The SANITAIRE Room Air Sanitizer is categorized as a germicidal ultraviolet fixture that utilize...

Page 5: ... The SANITAIRE Model RSCS280A Room Air Sanitizer includes 1 Model RSCS280A 120v 50 60Hz 1 4 High Output STER L RAY Germicidal Ultraviolet Lamps may be packed within the unit INSTALLATION Fixture Placement The SANITAIRE Model RSCS280A fixture is recommended for use in rooms with suspended T Bar ceilings Access to the room does not need to be controlled while the fixture is operating The exposure ch...

Page 6: ...onditions surrounding each installation It is recommended that the ultraviolet lamps be cleaned at least once every three 3 months The STER L RAY Germicidal Ultraviolet Lamps used in the SANITAIRE have a manufacturer s rated average effective life of 13 000 hours Lamps may operate longer than their rated effective life but the reduction in ultraviolet output will make it impractical to use past th...

Page 7: ...lamp by the base and carefully pull the lamp away from the connector while removing pressure off the STER L RAY lamp from the end supports See Figure 3 4 Once the lamp is clear from the connector carefully lift STER L RAY lamp away from the far end support and remove STER L RAY lamp from the unit See Figure 4 5 Repeat until all STER L RAY lamps have been removed from the unit Figure 4 Figure 3 6 M...

Page 8: ...fully push the STER L RAY lamp into the connector See Figure 7 12 Once the STER L RAY lamp is securely seated into the connector carefully push the STER L RAY lamp into the end support See Figure 8 13 Repeat until all STER L RAY lamps have been installed into the unit Avoid exposure to direct or reflected germicidal ultraviolet rays Germicidal ultraviolet rays are harmful to the eyes and skin 14 C...

Page 9: ...ater either by immersing or with a fine spray in the direction opposite to the airflow If the filter is contaminated with heavy grease concentrations wash filter with a detergent solution NOTE Follow all manufacturer s instructions and precautions for handling detergents The filter should be rinsed thoroughly with clean water 6 Shake off any excess water from PromateTM filter and let dry 7 Replace...

Page 10: ...nce Use personal protection equipment such long sleeves with no gaps between cuffs and gloves and ultraviolet resistant face shield if troubleshooting SANITAIRE IMPORTANT This unit is to be serviced ONLY by qualified and appropriately trained and or licensed personnel Table 1 Troubleshooting Problem Possible Cause Corrective Action Unit not operating No electrical power Verify that the unit is con...

Page 11: ...s for 4 Lamps 88W Weight Pounds lbs 64 lbs j Fixture capacity incorporates an estimated allowance for airflow friction loss across filter and fixture k Wattage is lamp watts only and does not include ballast loss l Ultraviolet output at 254nm at 100 hours and 80 F approximate All specifications dimensional data etc are approximate and subject to change without notice REPLACEMENT PARTS Table 3 Repl...

Page 12: ...l notify us at our main office in writing of claims including claims of breach of warranty within thirty 30 days after the Buyer discovers or should have discovered the facts upon which the claim is based Failure of the Buyer to give written notice of a claim within the time period shall be deemed to be a waiver of such claim The provisions of the above warranty are our sole obligation and exclude...

Reviews: