Installation of transducer and attentions
1. Mounting hole dimensions and sealing
2.The right and wrong sketch for installing holes
3. Attentions for installation
Installing holes' dimensions for 45
sealed slopes
Installing holes' dimensions for plain end face
Installing holes' dimensions for flange installation
Seal Surface
Seal Surface
Seal Surface
Fange
Diameter
Right
Wrong
Wrong
Wrong
1) Holes must be punched strictly in accordance with dimensions of installing holes for virgin
installation. Installing holes must be thoroughly cleared in after-installations;
2) Our company provides special tools for drilling and clean holes;
3) Test system should be inspected before installed to confirm that the measurement accuracy of
transducer/transmitter, the method described in the second table instructions;
4) Disassembly and installation should be conducted in hot state (material melting state).the
probe should be cleaned immediately when dismounting. Keep the probe clean. Be careful
with probe;
5) Conduct hot-line calibration with pressure gauge. When transducer/transmitter reaches full
thermal equilibrium after 1.5 hours installed, the details please see the calibration method of
second instrument. The PS series intelligent pressure gauge our company produced has the
automatic hot state calibration procedures, press SET button and CAL button and then loosen
CAL button until the pressure display zero pressure in thermal equilibrium to automatically
calibrate the test system;
6) Transducer and transmitter should be kept in dry environment when not used.