PRODUCT NUMBER: THDG986
Non-Contact Infrared Human
Forehead Thermometer Instructions
This is a non-contact infrared thermometer applicable to forehead measurement. The thermometer measures body temperature by collecting heat
radiation emitting from forehead. New probe structure is adopted in this thermometer. It’s a simply operated, hygienic, reliable and highly accurate
thermometer. Users can get precise reading within one second by one touch. This thermometer is cost-effective and is widely used in schools,
customs, hospitals and for family use.
This thermometer is also capable of measuring object temperatures ranging from 32°F–199.4°F (0°C–93°C).
This thermometer is classified as a Class II Medical Device, sorted as internally-powered equipment and type B application device. Water proof
rating is: IPXO. It is prohibited to use this thermometer in flammable anesthetic gas or gas mixture of air and oxygen or nitrous oxide. This is a
continuous operation equipment. It is classified as II a by EU.
Any object can generate a certain portion of infrared radiant energy per its own temperature. The radiant energy and its wavelength distribution
are subjected to its surface temperature. Based on this principle, this thermometer is designed to detect infrared radiation at 5 –14um by a highly
precise infrared sensor made in Germany. By adopting this high quality sensor plus special calculation and calibration, this thermometer is able to
take accurate body temperature.
• Adopting reliable sensor made in Germany
• All-new, patented probe design to ensure high accuracy
• Excellent adaption to ambient temperature. Accurate and reliable even under complicated surroundings
• Measurement mode for human body and object available, one button to switch measurement mode
• Beeper for fever or high temperature
• °C or °F reading available
• Automatic shut-off
• Storage of last 10 readings
• LCD with 3 backlight color options for easy reading
Effective Distance
5cm – 8cm
Measurement Method
Non-Contact
Range
Human body
95°F–109.2°F (35°C–42.9°C)
≤95°F/35°C: “Lo” Displays
>109.9°F/42.9°C: “Hi” Displays
Object
32°F–199.4°F (0°C– 93°C)
<32°F/0°C: “Lo” Displays
>199.4°F/93°C: “Hi” Displays
Accuracy
Human Body
±0.4°F/0.2°C
Object
±1.8°F/1.0°C
Resolution
0.1°F/0.1°C
Working Condition
60.8°F–104°F (16°C– 40°C)
RH≤85% Non-condensing
“ERR” displays when it is not used under working condition
Storage Condition
-4°F–131°F (-20°C –55°C)
RH≤85% Non-condensing
INTRODUCTION & CLASSIFICATION
FEATURES