20
Fig.14 Finish of calibration signal source
(5)
Set the path, click “Set Path”, and select the path of the DAC and temperature compensation file, and then
the follow-up measurement DAC file, S3302 test file as well as the compensation file will be placed into a folder
with the name of the spectrometer serial number.
(6)
Check S3302 series spectrum analyzer to ensure that there are no IF DAC files and compensation files. If a
DAC or compensation file exists, it shall be deleted firstly. The specific operation is as follows: Connect the PC
with the S3302 spectrum analyzer with a point-to-point network cable, open the S3302 host folder:
“ftp://172.141.xx.xx/” in PC browser, delete tc.tcdoc and IFGainAtten.dc in the MMCMemory folder, and then
restart the spectrum analyzer.
Attention: The IP address of the S3302 spectrometer and PC shall be located on the same network segment.
If they are different, the IP of the computer or spectrometer shall be set to make it within the same network
segment (for example, if the IP of the spectrometer is 172.141.11.202, the IP of the computer shall be set to
172.141.11.xx).
GPIB card
10MHz out
BNC cable
Signal source
Adapter 2.4 mm cable
Adapter
10MHz
in
Network cable
PC
Spectrum analyzer
Fig.15 Measure spectrum analyzer
Connect the computer with the spectrum analyzer with a point-to-point network cable; connect the BNC cable
with 10 MHz reference input of the spectrometer and 10 MHz reference output of the signal source, and then