43
S122
User /
Service Instructions
5 Configuration Mode
5.4 F1 - specification settings
Press
and
to
enter the required capacity:
Full ca 9d
max lb w
i
ght=(max kg
wghtx2.24)
60kgx20g 006018
150kgx50g 015045
300kgx100g 003009
130lbx0.05lb 013045
330lbx0.10lb 003309
660lbx0.20lb 066010
Press
and
to
enter the weight settings:
digit 1 = divisions (1, 2, or 5)
digit 2 = decimal places
(1, 2, or 3)
digits 3 - 6 = 0
ZERO
C00000
kg
lb
ENTER
013045
kg
130lb capacity
lb
ZERO
d00000
kg
lb
ENTER
530000
kg
lb
Divisions=0.05lb
ZERO