P14.x
0, 1, 2
0
Weight signal within power-on zero point range,
Choose which data as current power-on zero
point; 0= current weight ; 1= calibration
zero-point; 2=switch-off zero-point
P15.x
0, 1, 2
1
Weight signal not within power-on zero point
range, Choose which data as current power-on
zero point; 0= current weight ; 1= calibration
zero-point; 2=switch-off zero-point;
3=continuously display “
0ˉˉˉˉ”
P16.x
0, 1, 2, 3,
4, 5, 6,7,8
4
Zero tracking range: 0=0d, no tracking;
1=+0.25d; 2= +0.5d; 3=+1d; 4=+1.5d; 5=+2d;
6=+3d; 7= +4d; 8=+5d
P17.x
0, 1, 2, 3
2
Data filter intensity: 0=very weak, 1=weak,
2=middle, 3=strong
P18.x
0, 1….9
1
Check weight stability range: 0=+0.5d;
1
= ±
1d ; 2=+1.5d; 3=+2d; 4=+3d; 5=+4d ;
6=+5d ; 7=+6d ; 8=+7d; 9=+8d
P19.x
0, 1….9
3
Overload limit range: 0=FS+0d; 1=FS+9d;
2=101%FS; 3=102%FS; 4=105%FS;
5=110%FS; 6=120%FS; 7=150%FS;
8=200%FS; 9=No limitation
29