13
6.
Repeatedly press the
F
key to scroll through the choices. When your
choice is displayed, press the
TARE
key…
The display will briefly show 1/3 capacity, (10Lb for a 30 lb
capacity scale), and then display
0
. Press the
ZERO
key if the
display is not at
0
.
7.
Load weight equal to 1/3 of the scale capacity on the center of
platter, and press the
TARE
key when the display becomes stable…
2/3 capacity weight is displayed.
8.
Load weight equal to 2/3 of the scale capacity on the center of
platter, and press the
TARE
key when the display becomes stable…
Full capacity weight is displayed.
9.
Load weight equal to 100% of the scale capacity on the center of
platter, and press the
TARE
key when the display becomes stable…
The display shows
c 0
, with the
c
flashing. At this point you can
input a gravity adjustment.
10.
Press
ZERO
to increment the factor by one. Press and hold the
ZERO
key for 3 seconds to begin a rapid incrementing of the factor.
Press
ZERO
again to stop the automatic incrementing. Press
F
key
to decrease the factory by one. When you have the factor you want
displayed, press the
TARE
key to accept it…
CAEnd
(calibration end) will briefly display, then the unit returns
to normal weigh mode.
11.
Unload the weight, and turn off the scale. Turn on the scale again to
check the accuracy. If the scale is not accurate, please recalibrate
the scale as per above steps. If the scale is accurate, place
calibration switch to the off position, and replace the cover. If
needed, assemble the lead seal to lock the seal.