12
2.
Place the scale on stable, level surface and turn the unit on.
3.
After the self-check is complete, press and hold the
TARE
key
until…
CAL
is briefly displayed, then
0
.
4.
Empty the scale and press the
TARE
key to set the zero point.
Hg or Lb is displayed.
5.
Toggle between the choice of calibration weight unit of measure
using the
F
key. When the unit of measure you want to use is
displayed, press the
TARE
key…
The full capacity will flash.
6.
Load the full capacity weight onto the scale platform and press the
TARE
key.
CAEnd
(calibration end) will briefly display, then the unit returns
to normal weigh mode.
If you replace the PCB or loadcell, the scale requires recalibration. If the
display shows
Err-0
after power up, it means there is a problem with
calibration. Press and hold the
TARE
key to enter calibration and repeat
the calibration process.
Multipoint Calibration:
Follow these steps for multipoint calibration of the scale. You will need test
weights equal to 1/3, 2/3 and full capacity.
1.
Place the calibration switch, located on the bottom of the scale into
the ON position. You may need to remove the small cover to access
the calibration switch. Use a small screwdriver to remove the two
screws holding the cover.
2.
Place the scale on stable, level surface and turn the unit on.
3.
After the self-check is complete, press and hold the
F
and
ZERO
keys until…
20
or
100
is displayed. This stands for the powerup zero range.
4.
Repeatedly press the
F
key to toggle between the two choices:
20
and
100
. When your choice is displayed, press the
TARE
key…
Lb
or
Kg
is displayed. This is the calibration unit of measure.
5.
Press the
F
key to toggle between the two choices. When your
choice is displayed, press the
TARE
key…
Scale full capacity is displayed.