AI0-IF-025-00
2/9
* The product life of the instrument and the expected life of the main
components listed here is not the warranty period, but an average
number of years when the above conditions are met, so it will vary
depending on the operating environment and the method of use.
P. 1.7
Operating
Conditions
-
Add the following items.
Altitude: 2000 m or lower
Relative humidity 10 to 85% RH (non-condensing)
Electric shock protection: Class I
Overvoltage Category: II
Pollution degree: 2
Location of instrument use: Indoor use
Storage environment conditions: Ambient temperature -10 to 60°C
-
Replace Electrical Safety statement with the following.
For Model 6500 / 6503 –
cETL Certification
UL 61010-1, CAN/CSA-C22.2 No.61010-1, UL 61010-2-101,
CSA C22.2 No.61010-2-101
For Model 6501 –
Amendment of Pharmaceutical Affairs Law
JIS C 1010-1, JIS C 1010-2-101, JIS C1806-2-6, JIS T 14971,
ISO 13485, IEC 62366, IEC 62304
For Model 6502 –
CE Marking
Regulation (EU) 2017/746
EN 61010-1, EN 61010-2-101, EN 61326-2-6, EN 62366,
IEC 62304, EN ISO 18113-1, EN ISO 18113-3, EN ISO 14971
Directive 2011/65/EU + (EU)/2015/863
P. 2.1
Installation & setup
-
Add "Environment conditions for instrument installation."
A space allowing at least 40 cm on the left side of the instrument.
A space allowing at least 50 cm above the instrument.
A space allowing at least 40 cm in front of the instrument.
A space allowing at least 10 cm around the other side of the instrument.
(On the right side of the instrument, only when it is not connected to a
staining instrument.)
P. 2.2
Installation & setup
-
Add an option.
SG Guide 4(1) for SFE, product code N94-450-00