ENGLISH
36/37
Then follow the next steps:
- Press “
SEL
” for 3 seconds (“
dSP1 0040
” appears on the display).
- Press “
SEL
” again so that the display flashes.
- Press the down arrow until the value
0077
is set.
- Press “
SEL
” again so that the display stops flashing.
- Press the down arrow (“
dSP2
” appears on the display).
Repeat the operation 14 times with the values marked in the table below:
- Once the values have been modified, press “
SEL
” for 2 seconds to return to the temperature
display mode.
10.3.- Self-learning
Consult with the SAGOLA S.A.T. before carrying out any operation that
modifies the default parameters set in the Controller.
To carry out any modification you must make sure that the minimum
air consumption during the whole process must be at least 150 L/
min.
Symbol
Display
Factory default
Program
Locked
DSP1
DSP1
0040
0077
00FF
DSP2
DSP2
00E0
00E4
00E4
DSP3
DSP3
0060
00F0
00FE
DSP4
DSP4
002C
00A8
00FF
DSP5
DSP5
0014
009C
00FF
DSP6
DSP6
0001
00FF
00FF
DSP7
DSP7
0000
00FF
00FF
DSP8
DSP8
0000
00FF
00FF
DSP9
DSP9
0000
009F
00FF
DP10
DP10
004E
00FF
00FF
DP11
DP11
00bE
00FF
00FF
DP12
DP12
00FF
00FF
00FF
DP13
DP13
0007
014F
017F
DP14
DP14
0007
014F
07F