Press
<START/STOP>
to save or
<CURSOR>
and
<START/STOP>
to discard.
IMPORTANTE: Warm-up stabilization time.
The electronic components of the electrostatic sensor require a warm-up time of
about 5 minutes to obtain accurate and stable measurements.
A warning symbol lamp on the display and an acoustic signal advice that 5 minutes
have expired.
This symbol indicate
that warm-up is active
ZERO
During the warm-up it's possible to measure anyway, keeping in mind a little
inaccuracy.
If the unit does not indicate 0V while the probe is connected to ground and
WARM-
UP ALARM is no longer active
enter the ZERO procedure by pressing 2 times
<START/STOP>
at the screenshot of this MENU.
VIEW
CLEAR
COM
If there are graphs logged in the internal memory is possible to see them.
Press
<START/STOP>
: a pop-up window asks you which graph to view.
Move the cursor pressing <
CURSOR>
and select the graph pressing
<START/STOP>
Press
<LEFT>
or
<RIGHT>
to scroll the graph.
Press
<START/STOP>
again to exit.
If there are graphs logged in the internal memory is possible to delete them.
Press <
START/STOP>
: a pop-up window ask you what you want to delete (it's only
possible to delete the last graph or all the graphs memorized)
Move the cursor pressing
<CURSOR>
and select the command by pressing
<START/STOP>
In this menu the instrument can export data to a PC