background image

 

  

TECHNICAL FEATURES 

 

The MegaGuard Easy can measure surface resistance and resistance 
to the ground of mats, floors and other objects of the EPA area,

 

and can also be used for point-to-point measurements.

 

 

> 10

12   

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 T

 

 

οο

 

< 10

12 

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 G

 

- 1 T

 

< 10

11   

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 G

 

- 100 G

 

< 10

10   

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 G

 

- 10 G

 

< 10

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 M

 

- 1 G

 

< 10

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 M

 

- 100 M

 

< 10

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 M

 

- 10 M

 

< 10

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 K

 

- 1 M

 

< 10

5   

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 K

 

- 100 K

 

< 10

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-10K

 

 

Resolution: 1 decade 
Accuracy: ± 1 decade

 

Measuring voltage:  . . . . . . . . 10Vdc for R<100k

 

 

100Vdc for  R>100k

 

 

Dimensions:  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80x120x27mm

 

Weight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200g

 

Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 V

 

 

  

PROBES AND ACCESSORIES 

 

SG9265065  

. . . . . . . . . . . . . . . . . . Shielded probe made to IEC 61340-5-1

 

. . . . . . . . . . . . . . . . . . . . . . . . 63/30mm diameter, 2.5Kg weight (optional)

 

 

SG9265070  . . . . . . . . . . . 

Concentric ring PROBE made to IEC 61340-5-1

 

. . . . . . . . . . . . . . . . . . . . . . . . 63/30mm diameter, 2.5Kg weight (optional)

 

 

SG9265013B

. . . . . . . . . . . . . Connection cable jack 3.5mm - banana 4mm

 

Summary of Contents for MegaGuard easy

Page 1: ...SG9265055 Reference manual ...

Page 2: ...ator s security avoid to touch the electrodes directly WARRANTY This unit is guaranteed against all defects due to faulty materials and workmanship within 12 months from the date of purchase A use not conforming to what specified might be dangerous to the safety of the operator and may damage the instrument In such circumstances the manufacturer is relieved of any liability and the warranty itself...

Page 3: ...Ω 1 GΩ 10 8 10 MΩ 100 MΩ 10 7 1 MΩ 10 MΩ 10 6 100 KΩ 1 MΩ 10 5 10 KΩ 100 KΩ 10 4 0 10KΩ Resolution 1 decade Accuracy 1 decade Measuring voltage 10Vdc for R 100kΩ 100Vdc for R 100kΩ Dimensions 80x120x27mm Weight 200g Battery 9 V PROBES AND ACCESSORIES SG9265065 Shielded probe made to IEC 61340 5 1 63 30mm diameter 2 5Kg weight optional SG9265070 Concentric ring PROBE made to IEC 61340 5 1 63 30mm d...

Page 4: ...CONTENT OF THE PACKAGE The code SG9265055 includes 1 MegaGuard Easy with battery 1 connection cable SG9265013B 3 5mm jack banana4mm Manual ISO9000 traceable calibration certificate Box ...

Page 5: ...al probe RIGHT SOCKET Input for external probe 100V MEASURING VOLTAGE LOW BATTERY LED MEASUREMENT BUTTON BATTERY COVER BATTERY REPLACEMENT Open the battery cover unscrewing the 2 fixing screws Replace the battery Close the battery cover ...

Page 6: ...d to the 3 5mm jack Left Socket and Left Measuring Rubber In one of these two points the SIGNAL measurement input is present The SIGNAL measurement input is connected to the Left Measuring Rubber When the 3 5mm jack is inserted into the Left Socket the Left Measuring Rubber is automatically disconnected and the SIGNAL measurement input connected to the 3 5mm jack EXTERNAL PROBE CONNECTION SG926506...

Page 7: ... more after the value seems to be stable RESISTANCE TO GROUND MEASUREMENT Be sure that the surface under test is clean and free of grease or other chemicals than can create an isolant layer In case of doubt clean the surface with a specifical detergent without alcohol or silicon so that the non conductive layer will be removed Insert the 3 5mm jack of the connecting cable into the Left Socket SIGN...

Page 8: ...ts on it means that the applied test voltage is 100V When the LED is off the applied test voltage is 10V The displayed measure start always from the lower resistance valueand gradually reaches the real resistance value In order to get accurate measurements wait 5 seconds more after the value seems to be stable MEASURING PRECAUTIONS ATTENTION DURING GROUND RESISTANCE MEASURING if the earth referenc...

Reviews: