25
Dry
Dry the fiber rod with a lint-free cotton cloth.
Inspection and maintenance
After cleaning and disinfecting, the body light guide is visually inspected for cleanliness. No visible
contaminants are considered to be clean. If the body light guide is found to be corroded and rusted, stop
using it immediately.
package
Immediately after drying, place the body light guide in a steam sterilization bag for sealed packaging.
Attention
Steam sterilization bags in accordance with ISO 11607-1 should be used and the packaging
must be sealed with a sealing machine.
Sterilization
Use a high pressure steam sterilizer in accordance with EN 13060 for sterilization.
Sterilization in an autoclave according to ISO 17665-1.
1) sterilizable parts: body light guide;
2) Sterilization method: high pressure steam sterilization method;
3) Sterilization conditions: at 134 ° C, not less than 18 min.
Attention
The body light guide can be autoclaved, and no other parts are available.
Storage
Store the sterilization equipment in a dry, clean, dust-free environment at a suitable temperature of 10°C
to 40°C.
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION