1. Cleaning
■
Angle Handpiece
①
Rotate the head in a tap water about 10 seconds with maximum speed to remove the remains
(blood, physiologic saline solution, etc.) and rinse.
※
Do not put tap water in the motor.
②
Disconnect the angle handpiece to the motor.
③
Remove contaminants on the angle handpiece using a nylon brush after soaking for 20 minutes
in Enzymatic Detergent.
④
Rinse the angle handpiece using the tap water for 3 minutes to remove Enzymatic Detergent.
⑤
Wash the angle handpiece for 10 minutes in the ultrasonic cleaner containing Enzymatic Detergent.
⑥
Rinse the angle handpiece for 3 minutes in the ultrasonic cleaner containing purified water repeatedly
subjected 3 times. (Change the purified water each time.)
⑦
Wipe using dust-absorbent after the product soaked in alcohol(70%) for 30 seconds.
■
Motor
Wipe the surface of motor with alcohol swab or paper towel after removing the dirt.
※
Pay attention to not enter the water into the motor.
MAINTENANCE OF THE PRODUCT
- Proteolytic enzyme detergents are recommended as Enzymatic Detergent.
- It is recommended to use ethanol for alcohol (70%).
30
2. Sterilization
Sterilization Product List
· Angle handpiece
· Motor (Assembling with motor cap)
Sterilization Method
①
Dry the applicable products completely after cleaning.
②
Put the applicable products in a sterile bag and seal tightly.
※
Make sure to assemble the motor cap for autoclave on the motor.
※
Do not disassemble the motor and motor cord when sterilizing.
③
Put the sterile bag in the sterilizer and sterilization for 4 minutes with 132
℃
, and let it dry for 30 minutes.
Temperature
132
℃
Time
4 minutes
Dry
30 minutes
Sterilization
Moist heat sterilization
- This product is designed to withstand 250 times of re-sterilization process.
- This product recommends the moist heat sterilization (autoclave), other sterile methods are
not guaranteed the effectiveness.
- Make sure to use the sterilization bag and seal tightly.
- Make sure to dry the applicable products before sterilization.
- Motor could be damaged in sterilization because of the moisture inflow or else, so should be careful to handle it.
- Sterilizing the angle handpiece with blood or other remains may cause the breakage of the angle handpiece.
Therefore, it is necessary to clean the angle handpiece before sterilization.
- Irrigation tube and Y-tube are disposable. Do not re-use.
- Keep dry a connector and socket of the control box.
- Before using, dry the product completely.
132