2. Components

 : Airpiece, Repair Kit (Wrench, Cleaning Pin, Brush), 
   Spray Nozzle, Instruction Manual,
   Spare O-Ring (CAB10LN/  CAB10LNTi)

1. Intended Use

 : A handpiece driven by air pressure to remove teeth and etc.

3. How to connect each part

Please read the instruction manual carefully before operation and keep it for future reference.

AIRPIECE ( ATN400, CAB10LN, CAB10LNTi, CAB10LK )

3.1 Airpiece

 1) 

Mounting (TRAUS ATN400 / TRAUS CAB10LN / TRAUS CAB10LNTi)

  ① Insert the coupling into the hose and tighten the hose nut.
  ② Insert the air piece into the coupling.
  ③ Connect the air piece and the coupling by pulling connection ring 
  ④ Grasp the coupling and connect the air piece and the coupling to the end.
  ⑤ Check if the air piece is pulled out or not.

 2) Isolation

① Pull the connection ring of the coupling and remove the air piece.
   (ATN400 / CAB10LN / CAB10LNTi)
② Remove the airpiece while holding the coupling and pulling the airpiece. (CAB10LK)

3.2 Bur
 1) Mounting

  ① Insert the bur
  ② Press the button on the airpiece head and
      connect the bur to the end.
  ③ Release the button
  ④ Make sure that the bur is not pulled out 
      by pulling out

 2) Isolation

  ① Pull out the bur pressing the button on the airpiece head

Warranty Exclusions

4. Maintenance

4.1 Cleaning

 ① Rotate the head in a tap water about 10 seconds with maximum speed to
     remove the remains (blood, physiologic saline solution, etc.) and rinse.
 ② Remove contaminants on the airpiece using a nylon brush after soaking
     for 20 minutes in Enzymatic Detergent.
 ③ Rinse the airpiece using the tap water for 3 minutes to remove Enzymatic Detergent.
 ④ Wash the airpiece for 10 minutes in the ultrasonic cleaner containing Enzymatic Detergent.
 ⑤ Rinse the airpiece for 3 minutes in the ultrasonic cleaner containing purified

     

water repeatedly subjected 3 times. (Change the purified water each time.)

 ⑥ Wipe using dust-absorbent after the product soaked in alcohol(70%) for 30 seconds.

4.2 Lubrication

 

① Insert the spray nozzle to hose connection part(the end part of airpiece)

     and spray lubricant 2~3 seconds.

4.4 Replace the air turbine cartridge

  ① Mount the bur on the chuck.

  ② Loosen the head cap with the head cap wrench.

  ③ Push the bur from the head to remove the cartridge.

  ④ Clean the inside of the head.

  ⑤ Put the Cartridge slowly into the head. Be careful not to damage the ball bearing.

  ⑥ Lock the head cap with the head cap wrench.

4.5 O-ring exchange <TRAUS CAB10LN/TRAUS CAB10LNTi only>

  Replace the O-ring if you notice any water coming out from the exhaust.

  If left un-replaced, water may leak from the coupling connection.

  Be sure to replace as a set with the spare o-ring of the component.

  ① Remove the handpiece from the coupling.

  ② Remove the 'Handle B'.

  ③ Take out the old O-ring with your fingers and insert the spare O-ring into

      the groove. Be careful not to damage the O-ring.

  ④ Tighten the 'Handle B' firmly to the handpiece.

    ※ 1 set of O-ring consists of 5 O-rings.

4.6 Cleaning Glass Rod

  

If dirt or foreign material adheres to the glass rod, wipe it with a cotton swab

  dipped in sterilized alcohol. 

5. Warranty

  SAESHIN guarantees this product for 1 year from invoice date. Product warranty

  

means that SAESHIN has responsibility for defective material or operation failure.

  

The product warranty does not cover user’s misuse, wrong installation, inappropriate

  maintenance and repair, and normal wear of consumable parts such as cartridge
  and bearing. In order for SAESHIN to confirm if the warranty is valid will require
  checking the operating condition, environment information, serial number and

  

stamped invoice. The warranty will be performed as a repair or exchange according

  to manufacturer’s analysis of product and judgement.

Misuse and improper handling by users
Dropping the products during installation, movement or use.
If the unit is damaged due to unspecified consumable parts or accessories.
If the product is not repaired by Saeshin or the repair shop designated by Saeshin.
Normal wear of consumable parts.
Damage and failure due to natural disaster.
If the product is disassembled or modified at will.

- Use of any other purpose is prohibited.
- The company did not sterilize the product. Please make sure to sterilize before
  operation. (Please refer to the 4.3 Sterilization)
- Do not drop the product or force with strong external shocks.
- Do not disassemble or modify the product.
- Check for vibration, heat, noise and other abnormal symptoms before use.
- If the product operates abnormally, stop immediately and contact the seller for check.
- Do not remove the air hose while the product is rotating.
- Moisture and impurities may cause malfunction of the product. 
  Please clean and dry with air.
- Make patient safety the top priority when use.
- To maintain it clean, always wipe the product with rubbing alcohol before use.
- Improper use may cause damage to the product. Be sure to use it by a skilled
  technician or specialist.

The manufacturer is not responsible for accidents caused by use not in accordance

  with the instruction manual.
- The user is responsible for the proper use and maintenance of the product.
- This product should not be disposed with general waste.

※  If the spray is not working well,
    clean the spray port using the
    cleaning pin as shown below.

※  If there is foreign material around
    the clean head hole and it cannot

    

be removed, use a brush to remove it.

※ The rotation speed varies slightly depending on the direction of rotation
   and the shape of the hose used.
※ Coupling: Designed to be compatible with NSK Coupling.
   ( ATN400, CAB10LN, CABLNTi)
※ Coupling: Designed to be compatible with KAVO Coupling. ( CAB10LK)

Weight

RPM

Bur Type

Max. Bur Length

Max. Working

Part Diameter

Recommended

Drive Air Pressure

- Lubricate the product after each use.
- When lubrication, please hold the product firmly because
  the product can go out by the spray pressure.
- The nozzle tip is not included in a set.

Be sure to check if the product is cleaned, lubricated and sterilized

  before using it.

Remove the bur and disassemble the airpiece after checking if the

  operation stopped for sure.

Do not use the bur that exceed the maximum length. The maximum

  length of the bur is 21mm.

When using the bur, follow the instructions provided by the bur manufacturer.

-

 Do not clean using thinner, benzene, or other solvents.

- Do not forcibly insert the cleaning pin into the spray port or push it tilted.
  The injection nozzle may be deformed, which may cause malfunction.
- After washing, perform a visual inspection in a place with sufficient light.
  If contamination is found on the surface, please re-clean according to
  our cleaning method.

- Replace the cartridge when you feel abnormal vibration, noise or
  heat while test or operation.

When cleaning the glass rod, sharp edges may cause scratches on its

  surface, which may reduce the light transmittance.

[Precautions]

Use conditions

Temperature: 10~35℃

Humidity: 10~80%

Atmospheric pressure: 700~1,060hPa

Shipping/Storage conditions

Temperature: -20~60℃

Humidity: 10~90%

Atmospheric pressure: 500~1,060hPa

400,000 rpm ± 10% (At the recommended drive air pressure)

ISO 1797 Ø1.59 ~ Ø1.60mm Standard Bur

21mm

Ø2mm

0.25 ~ 0.35Mpa (2.5 ~ 3.5kgf/cm2)

<66NL/min (0.3Mpa)

0.08 ~ 0.20Mpa (0.8 ~ 2.0kgf/cm2)
0.10 ~ 0.20Mpa (1.0 ~ 2.0kgf/cm2)

0.30Mpa (2.8kgf/cm2)

TRAUS CP10NQ

(Compatible with NSK QD-J M4)

Glass Rod

TRAUS CP10LN

(Compatible with NSK PTL-CL-LED)

Glass Rod

TRAUS CP10LN

(Compatible with NSK PTL-CL-LED)

-

Model

Picture

Head

Cartridge Head Cap Wrench

※ If the chuck is contaminated, it may cause problems in tightening.
   Please lubricate the chuck as shown below once a week.

Nozzle

Tip

본 제품은 

의료기기

 입니다. / This product is 

medical device.

4.3 Sterilization

 

① Dry the product completely after cleaning.                                        

 ② Put the product in a sterile bag and seal tightly.
 ③ Put the sterile bag in the sterilizer and sterilization for 4 minutes with 132℃,
     and let it dry for 30 minutes.

- Maximum sterilization number is 250 times.
- Make sure to use the sterilization bag and seal tightly.
- Make sure to dry the applicable products before sterilization.

Sterilizing the product with blood or other remains may cause the

  

breakage of the product. Therefore, it is necessary to clean the product

  before sterilization.

Temperature

132℃

Time

4 minutes

Dry Time

30 minutes

Sterilization

Moist heat sterilization

(pre-vaccum)

Operating Drive

Air Pressure Range

Max. Air Consumption

Water Pressure

Chip Air Pressure

Optic

Coupling

Spray

Nozzle

Spray

Nozzle

<ATN400>

<CAB10LN / CAB10LNTi>

Hose Nut

Hose Nut

Airpiece

Coupling

Connection Ring

Connector Ring

Connection

Airpiece

<TRAUS ATN400>

<TRAUS CAB10LN/TRAUS CAB10LNTi>

Clean Head Hole

Clean Head

Hole

Brush

Spray Port

Cleaning Pin

Spray Port

53g

58g

45g

TRAUS ATN400

TRAUS CAB10LN

TRAUS CAB10LNTi

Head Button

Handle B

O-Ring

Glass Rod

Cotton Swab

52, Secheon-ro 1-gil, Dasa-eup,
Dalseong-gun, Daegu, Korea
Tel. 82-53-587-2341 / Fax. 82-53-580-0999

www.saeshin.com / [email protected]

대구광역시 달성군 다사읍 세천로
1길 52 (성서5차산업단지)

MERIDIUS MEDICAL EUROPE LIMITED

Unit 3D, North Point House North point
Business Park New Mallow Road CORK,
T23AT2P, Ireland
+353 212066448

1639

NON

STERILE

SOM-S077P-KE   Rev.7  2021-01-13

Connection

<TRAUS CAB10LK>

Hose Nut

Coupling

Airpiece

Airpiece

65g

TRAUS CAB10LK

Spray Nozzle

<CAN10LK>